iSite PACS image management system; Picture Archiving and Communications System; Radiology software package...
Philips Healthcare Informatics, Inc. began a recall of iSite PACS image management system; Picture Archiving and Communications System; Radiology software package, Philips Healthcare Informatics, Foster City, CA... on . The reason FDA records is "Incorrect display of one of the patient's images. This malfunction could cause an image to be incorrectly displayed, for example flipped left to right and anterior to posterior. The original image is still presented in...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1141-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- iSite PACS image management system; Picture Archiving and Communications System; Radiology software package, Philips Healthcare Informatics, Foster City, CA 94404.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Incorrect display of one of the patient's images. This malfunction could cause an image to be incorrectly displayed, for example flipped left to right and anterior to posterior. The original image is still presented in the correct orientation next to the incorrect image when displayed on the monitor.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips Healthcare Informatics mailed customers notification: Urgent Product Correction regarding FCO#2954704-01/04/08-001-C. Firm released a corrected version of software (iSite PACS v.3.5.69) on January 21, 2008. Firm is scheduling customers to upgrade to the corrected version.
- Quantity in commerce
- 197 units
- Distribution
- Worldwide Distribution: USA, United Arab Emirates, Germany, the Netherlands, Denmark, Australia, France, Italy, Belgium, Ireland, Italy, and Canada.
- Product codes and lots
- iSITE PACS version 3.5x (Versions 3.5 and up to and including 3.5.65): 3.5.0.27, 3.5.0.36, 3.5.36.4, 3.5.38, 3.5.51.0, 3.5.51.02, 3.5.51.05, 3.5.56.0, 3.5.57.1, 3.5.57.2, 3.5.57.3, 3.5.57.4, and 3.5.65.0