Z-0791-2008
CORDIS "Fire Star" 2.25 x 10 Dilatation Catheter, Catalog # 80110225, Distributed by Cordis Corporation...
Cordis US Corp. / initiated 2008-01-14 / Terminated / root cause: manufacturing process
Cordis Corporation began a recall of CORDIS "Fire Star" 2.25 x 10 Dilatation Catheter, Catalog # 80110225, Distributed by Cordis Corporation, Miami Lakes, FL 33014 on . The reason FDA records is "Slow Deflation or No Deflation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0791-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cordis US Corp.
- Product
- CORDIS "Fire Star" 2.25 x 10 Dilatation Catheter, Catalog # 80110225, Distributed by Cordis Corporation, Miami Lakes, FL 33014
- Product code
- LOX Catheters, Transluminal Coronary Angioplasty, Percutaneous
- Reason for recall
- Slow Deflation or No Deflation
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Cordis initated the recall on 1/14/2008. For U.S. customers: An Urgent Medical Device Recall letter and an Acknowledgement Form was sent to the following 4 individuals in each account: Cath Lab Director, Operating Room Director, Materials Director, Risk Manager; to be followed up with direct Cordis representative contact to facilitate obtaining signature, faxing form, collecting and returning units. For non-US customers, email notification with return acknowledgement is used to notify the affiliate distributors, who then notify and reconcile the customers in those countries. All users were instructed to immediately set aside product and return it for replacement.
- Quantity in commerce
- 133,300 units total for recalls Z-0747-0824-2008
- Distribution
- Worldwide.
- Product codes and lots
- Lot #s: 13188606 13206122 13218314 13223382 13235878 13240535 13246853 13253243 13260781 13276280 13290336 13293672 13296511 and 13296512
FIRST ARTICLE / WEEKLY