ev3 IntraStent LD Max Large Diameter Biliary Stent, Model Number: S18-16, ev3, 4600 Nathan Lane North...
Ev3, Inc began a recall of ev3 IntraStent LD Max Large Diameter Biliary Stent, Model Number: S18-16, ev3, 4600 Nathan Lane North, Plymouth, MN 55442, Catalog Number: 90-2319-000. ev3... on . The reason FDA records is "Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1129-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- ev3 IntraStent LD Max Large Diameter Biliary Stent, Model Number: S18-16, ev3, 4600 Nathan Lane North, Plymouth, MN 55442, Catalog Number: 90-2319-000. ev3 IntraStent LD Max Large Diameter Peripheral Stent Model Number: S18-16, Catalog Number: 90-2337-000 (not approved within USA).
- Product code
- FGE Stents, Drains And Dilators For The Biliary Ducts
- Reason for recall
- Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- ev3 notified all consignees via "Medical Device Recall" letter on 1/18/08. The letter was addressed to the Hospital and it described the device affected and the problem. The letter provided recommendations to segregate the affected devices and requested the return of the "Device Recall Field Action Reconciliation Form."
- Quantity in commerce
- 12
- Distribution
- Worldwide Distribution - USA, ARGENTINA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, DENMARK, GERMANY, GREECE, HONG KONG, IRAN, IRELAND, ITALY, JORDAN, KUWAIT, LEBANON, NORWAY, POLAND, RUSSIA, SAUDI ARABIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TURKEY, UNITED ARAB EMIRATES, and UNITED KINGDOM.
- Product codes and lots
- Lot Numbers: 3766729 and 2417395