Z-1042-2008
Siemens Magnetom Harmony Mobile MRI System. Model Number: 4759309, Siemens Medical Solutions USA, Malvern PA...
Siemens Medical Solutions USA, Inc. / initiated 2007-12-01 / Terminated / root cause: design
Siemens Medical Solutions USA, Inc began a recall of Siemens Magnetom Harmony Mobile MRI System. Model Number: 4759309, Siemens Medical Solutions USA, Malvern PA, 19355 on . The reason FDA records is "Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1042-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Magnetom Harmony Mobile MRI System. Model Number: 4759309, Siemens Medical Solutions USA, Malvern PA, 19355
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm issued an Update Instruction package to consignees 12/07. As an immediate measure, the recalling firm is initiating field checks of mobile MRI systems per Update Instruction MR039/07/S. This update defines a safety inspection for mobile systems, including checks for potential ice formation and proper configuration of venting assemblies.
- Quantity in commerce
- 12 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Serial Numbers: 10018, 10022, 10029, 10032 10036, 10160, 10167, 10255, 10289, 10316, 10386, and 10429.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
