Siemens Magentom Symphony, Model Number: 4760265, Siemens Medical Solutions USA, Malvern PA, 19355

Siemens Medical Solutions USA, Inc. / initiated 2007-12-01 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Siemens Magentom Symphony, Model Number: 4760265, Siemens Medical Solutions USA, Malvern PA, 19355 on . The reason FDA records is "Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1045-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Magentom Symphony, Model Number: 4760265, Siemens Medical Solutions USA, Malvern PA, 19355
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.
Root cause tag
design
FDA root cause
Device Design
Action
The firm issued an Update Instruction package to consignees 12/07. As an immediate measure, the recalling firm is initiating field checks of mobile MRI systems per Update Instruction MR039/07/S. This update defines a safety inspection for mobile systems, including checks for potential ice formation and proper configuration of venting assemblies.
Quantity in commerce
2 units
Distribution
Nationwide Distribution
Product codes and lots
Serial Numbers: 13309 and 13390