Z-1047-2008
Siemens Magnetom Symphony syngo, Model Number: 7104594, Siemens Medical Solutions USA, Malvern PA, 19355
Siemens Medical Solutions USA, Inc. / initiated 2007-12-01 / Terminated / root cause: design
Siemens Medical Solutions USA, Inc began a recall of Siemens Magnetom Symphony syngo, Model Number: 7104594, Siemens Medical Solutions USA, Malvern PA, 19355 on . The reason FDA records is "Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1047-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Magnetom Symphony syngo, Model Number: 7104594, Siemens Medical Solutions USA, Malvern PA, 19355
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm issued an Update Instruction package to consignees 12/07. As an immediate measure, the recalling firm is initiating field checks of mobile MRI systems per Update Instruction MR039/07/S. This update defines a safety inspection for mobile systems, including checks for potential ice formation and proper configuration of venting assemblies.
- Quantity in commerce
- 141 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Serial Numbers: 22021, 22024, 22060, 22074, 22079, 22108, 22115, 22116, 22121, 22127, 22172, 22196, 22225, 22242, 22272, 22306, 22311, 22342, 22352, 22353, 22356, 22402, 22440, 22441, 22446, 22471, 22498, 22510, 22520, 22523, 22530, 22561, 22565, 22596, 22598, 22604, 22619, 22621 22622, 22636, 22674, 22690, 22698, 22711, 22801, 22810, 22825, 22835, 22836, 22865, 22881, 22885, 22907, 22911, 22914, 22916, 22944, 22967, 23001, 23057, 23059, 23085, 23099, 23104, 23132, 23136, 23156, 23176, 23199, 23218, 23247, 23252, 23255, 23284, 23315, 23323, 23357, 23381, 23491, 23497, 23507, 23538, 23552, 23574, 23578, 23579, 23580, 23585, 23590, 23594, 23596, 23602, 23604, 23605, 23610, 23613, 23628, 23641, 23642, 23672, 23682, 23687, 23691, 23698, 23731, 23760, 23762, 23764, 23782, 23783, 23788, 23796, 23804, 23823, 23834, 23839, 23883, 23885, 23902, 23904, 23913, 23919, 23932, 23943, 23950, 23956, 23984, 23986, 23987, 23994, 23997, 23999, 24017, 24019, 24022, 24030, 24043, 24071, 24099, 24104, and 24107.