Z-1048-2008
Siemens Magentom Symphony syngo Upgrade, Model Number: 7106557, Siemens Medical Solutions USA, Malvern PA...
Siemens Medical Solutions USA, Inc. / initiated 2007-12-01 / Terminated / root cause: design
Siemens Medical Solutions USA, Inc began a recall of Siemens Magentom Symphony syngo Upgrade, Model Number: 7106557, Siemens Medical Solutions USA, Malvern PA, 19355 on . The reason FDA records is "Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1048-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Magentom Symphony syngo Upgrade, Model Number: 7106557, Siemens Medical Solutions USA, Malvern PA, 19355
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm issued an Update Instruction package to consignees 12/07. As an immediate measure, the recalling firm is initiating field checks of mobile MRI systems per Update Instruction MR039/07/S. This update defines a safety inspection for mobile systems, including checks for potential ice formation and proper configuration of venting assemblies.
- Quantity in commerce
- 35 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Serial Numbers: 14001, 14087, 14101, 14182, 14184, 14186, 14206, 14220, 14235, 14236, 14238, 14247, 14252, 14264, 14272, 14276, 14278, 14312, 14334, 14344, 14357, 14359, 14360, 14362, 14268, 14427, 14424, 14427, 14430, 14431, 14434, 14436, 14438, 14443, and 14447.