Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1287-2008

Edwards LifeStent FlexStar Biliary Stent System; Biliary catheter and accessories; Model Number: EX090303CD...

Edwards Lifesciences LLC / initiated 2007-12-05 / Terminated / root cause: packaging

Edwards Lifesciences, Llc began a recall of Edwards LifeStent FlexStar Biliary Stent System; Biliary catheter and accessories; Model Number: EX090303CD; Edwards Lifesciences LLC, Irvine, CA 92614 on . The reason FDA records is "Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1287-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
Edwards LifeStent FlexStar Biliary Stent System; Biliary catheter and accessories; Model Number: EX090303CD; Edwards Lifesciences LLC, Irvine, CA 92614.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Consignees were notified by an Urgent Product Recall letter sent on 12/5/07. The letter instructed users to return any affected product to the recalling firm for replacement products. The letter also advised users that the likelihood of patient complications is extremely low for products that have already been implanted in patients. Users were requested to return a confirmation form after checking the inventory for any affected products. For additional information, contact 1-949-250-3779.
Quantity in commerce
1 UNIT
Distribution
Nationwide Distribution
Product codes and lots
Lot Number: FR6H0184