Z-0956-2008
Medtronic SynchroMed EL Programmable Pumps, Model 8627-18, Medtronic, Inc., 710 Medtronic Parkway...
Medtronic Neuromodulation / initiated 2007-08-03 / Terminated / root cause: design
Medtronic Neuromodulation began a recall of Medtronic SynchroMed EL Programmable Pumps, Model 8627-18, Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. (The implantable Medtronic... on . The reason FDA records is "Pump Motor Stall; pumps can stall due to gear shaft wear". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0956-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Neuromodulation
- Product
- Medtronic SynchroMed EL Programmable Pumps, Model 8627-18, Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. (The implantable Medtronic SynchroMed EL Programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site)
- Product code
- LKK Pump, Infusion, Implanted, Programmable
- Reason for recall
- Pump Motor Stall; pumps can stall due to gear shaft wear.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Begining August 3, 2007, consignees were notified by letter. The letter "Medtronic Urgent: Medical Device Correction Aug 2007" described the problem and the Affected Devices as well as Patient Risk, Patient Management Recommendations, Next Steps and Physician and Patient Support.
- Quantity in commerce
- 81,023 total for recalls Z-0950-0957-2008
- Distribution
- Worldwide-USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, DS, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and countries of HONG KONG, JAPAN, AUSTRALIA, CANADA, BELGIUM, CROATIA, AUSTRIA, CYPRUS, The CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, IRAN, ICELAND, IRELAND, ITALY, JORDAN, KWAIT, LEBANON, MALTA, The NETHERLANDS, NORWAY, POLAND, PORTUGAL, SAN MARINO, SAUDI ARABIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TUNISIA, TURKEY, UNITED ARAB EMIRATES, The UNITED KINGDOM, CHILE, VENEZUELA, BRAZIL, and MEXICO.
- Product codes and lots
- See codes for all models in associated Recall Z-0950-2008