Medtronic Sofamor Danek USA, CD Horizon Spinal System Agile Dynamic Stabilization Device, REF 7881070, 10mm...

Medtronic Sofamor Danek USA Inc / initiated 2007-11-12 / Terminated / root cause: design

Medtronic Sofamor Danek USA Inc began a recall of Medtronic Sofamor Danek USA, CD Horizon Spinal System Agile Dynamic Stabilization Device, REF 7881070, 10mm, 5.5mm dia x 70mm, size: 10mm x 70mm, not for... on . The reason FDA records is "Breaks: Shear failure of the cable component of the system". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1430-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Sofamor Danek USA Inc
Product
Medtronic Sofamor Danek USA, CD Horizon Spinal System Agile Dynamic Stabilization Device, REF 7881070, 10mm, 5.5mm dia x 70mm, size: 10mm x 70mm, not for distribution in the U.S.A. or its territories, for spinal fixation.
Product code
NQP Posterior Metal/Polymer Spinal System, Fusion
Reason for recall
Breaks: Shear failure of the cable component of the system.
Root cause tag
design
FDA root cause
Device Design
Action
Medtronic Sofamor Danek initiated its recall on 12/12/2007 by letter to all Risk Managers who received the product at their hospital. Letter were also mailed to surgeons who have implanted the product. Sales Representatives were notified via a conference call and letter to follow-up with thier respective surgeons who have implanted the product to verify that their surgeon received the communication of the recall and that they are aware that patient monitoring is recommended.
Quantity in commerce
217 units
Distribution
Worldwide Distribution: USA, Canada, India, Venezuela, Netherlands, Chile, Brazil, and Australia.
Product codes and lots
ALL CODES