Z-1052-2008

Draeger GT500. GlobeTrotter Transport System. Catalog Number: MU20508, Draeger Medical, Inc., Telford, PA...

Draeger, Inc. / initiated 2008-01-31 / Terminated / root cause: manufacturing process

Draeger Medical, Inc. began a recall of Draeger GT500. GlobeTrotter Transport System. Catalog Number: MU20508, Draeger Medical, Inc., Telford, PA 18969 on . The reason FDA records is "Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1052-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger, Inc.
Product
Draeger GT500. GlobeTrotter Transport System. Catalog Number: MU20508, Draeger Medical, Inc., Telford, PA 18969
Product code
FMZ Incubator, Neonatal
Reason for recall
Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or [email protected].
Quantity in commerce
2 units
Distribution
Worldwide Distribution-- USA including states MO, OH, and PR, and countries of Australia, Canada, Columbia, Israel, UAE, France, China, Italy, Denmark, Switzerland, and UK.
Product codes and lots
Serial Numbers: PK00983 and PK00982