Draeger GT500. GlobeTrotter Transport System. Catalog Number: MU20508, Draeger Medical, Inc., Telford, PA...
Draeger Medical, Inc. began a recall of Draeger GT500. GlobeTrotter Transport System. Catalog Number: MU20508, Draeger Medical, Inc., Telford, PA 18969 on . The reason FDA records is "Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1052-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger, Inc.
- Product
- Draeger GT500. GlobeTrotter Transport System. Catalog Number: MU20508, Draeger Medical, Inc., Telford, PA 18969
- Product code
- FMZ Incubator, Neonatal
- Reason for recall
- Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or [email protected].
- Quantity in commerce
- 2 units
- Distribution
- Worldwide Distribution-- USA including states MO, OH, and PR, and countries of Australia, Canada, Columbia, Israel, UAE, France, China, Italy, Denmark, Switzerland, and UK.
- Product codes and lots
- Serial Numbers: PK00983 and PK00982