Trident Hemispherical Cluster; 42mm - 74 mm Arc Deposited; Hydroxylapatite Coated; Multiholed; Use with Size...
Stryker Howmedica Osteonics Corp. began a recall of Trident Hemispherical Cluster; 42mm - 74 mm Arc Deposited; Hydroxylapatite Coated; Multiholed; Use with Size A insert; hip prosthesis component, Stryker... on . The reason FDA records is "Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1170-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Trident Hemispherical Cluster; 42mm - 74 mm Arc Deposited; Hydroxylapatite Coated; Multiholed; Use with Size A insert; hip prosthesis component, Stryker Orthopaedics, Howmedica Osteonics Corp, Mahwah, NJ 07430
- Product code
- MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
- Reason for recall
- Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Recall notification letters were sent to Stryker Branches/agencies, OR Supervisors and Chief of Orthopaedics on 1/21/08. A Patient information sheet was sent on 2/4/08 to surgeons and hospitals. Per call with Center 2/4/08, recall letter to be revised and RES updated upon receipt. The letter was revised and sent on February 28, 2008 to include Trident PSL Acetabular shells. This letter was sent to Risk Management at hospitals and included the scope of the recall , the potential hazard and recommendation to physicians to monitor patients consistent with care for those receiving total hip replacement. It was also sent to surgeons and a revised letter to Stryker branches. A product acknowledgement form was included in all letters to indicate receipt of letter and quantity of product on hand, if applicable.
- Quantity in commerce
- 129,312 total shells, all Trident Hemispherical types
- Distribution
- Nationwide distribution.
- Product codes and lots
- Catalog Numbers: # 502-01-42A; 42mm; 502-01-44B, 44mm; 502-01-46C, 46mm; 502-01-48D, 48mm; 502-01-50D, 50mm; 502-01-52E, 52mm; 502-01-54E, 54mm; 502-01-56F, 56mm; 502-01-58F, 58mm;, 502-01-60G, 60mm; 502-01-62G, 62mm; 502-01-64H, 64mm; 502-01-66H, 66mm; 502-01-681, 68mm; 502-01-701, 70mm;, 502-01-72J, 72mm; 502-01-74J, 74mm. All lot codes with expiration dates between 1/2005 and 12/2012