Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1170-2008

Trident Hemispherical Cluster; 42mm - 74 mm Arc Deposited; Hydroxylapatite Coated; Multiholed; Use with Size...

Howmedica Osteonics Corp / initiated 2008-01-21 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Trident Hemispherical Cluster; 42mm - 74 mm Arc Deposited; Hydroxylapatite Coated; Multiholed; Use with Size A insert; hip prosthesis component, Stryker... on . The reason FDA records is "Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1170-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Trident Hemispherical Cluster; 42mm - 74 mm Arc Deposited; Hydroxylapatite Coated; Multiholed; Use with Size A insert; hip prosthesis component, Stryker Orthopaedics, Howmedica Osteonics Corp, Mahwah, NJ 07430
Product code
MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Reason for recall
Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Recall notification letters were sent to Stryker Branches/agencies, OR Supervisors and Chief of Orthopaedics on 1/21/08. A Patient information sheet was sent on 2/4/08 to surgeons and hospitals. Per call with Center 2/4/08, recall letter to be revised and RES updated upon receipt. The letter was revised and sent on February 28, 2008 to include Trident PSL Acetabular shells. This letter was sent to Risk Management at hospitals and included the scope of the recall , the potential hazard and recommendation to physicians to monitor patients consistent with care for those receiving total hip replacement. It was also sent to surgeons and a revised letter to Stryker branches. A product acknowledgement form was included in all letters to indicate receipt of letter and quantity of product on hand, if applicable.
Quantity in commerce
129,312 total shells, all Trident Hemispherical types
Distribution
Nationwide distribution.
Product codes and lots
Catalog Numbers: # 502-01-42A; 42mm; 502-01-44B, 44mm; 502-01-46C, 46mm; 502-01-48D, 48mm; 502-01-50D, 50mm; 502-01-52E, 52mm; 502-01-54E, 54mm; 502-01-56F, 56mm; 502-01-58F, 58mm;, 502-01-60G, 60mm; 502-01-62G, 62mm; 502-01-64H, 64mm; 502-01-66H, 66mm; 502-01-681, 68mm; 502-01-701, 70mm;, 502-01-72J, 72mm; 502-01-74J, 74mm.   All lot codes with expiration dates between 1/2005 and 12/2012