Stryker COCR V40 Femoral Head; xx-Long (+16mm) Neck Length 28mm Diameter Sterile, Forged Vitallium Alloy...
Stryker Howmedica Osteonics Corp. began a recall of Stryker COCR V40 Femoral Head; xx-Long (+16mm) Neck Length 28mm Diameter Sterile, Forged Vitallium Alloy; Product Number: 6260-9-528; Howmedica Osteonics Corp... on . The reason FDA records is "Fracture of Stem Neck - The firm recently performed laboratory testing indicates that there may be a potential increased risk of stem neck fractures when a V40 +16mm offset femoral head is mated with certain Accolade...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0963-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker COCR V40 Femoral Head; xx-Long (+16mm) Neck Length 28mm Diameter Sterile, Forged Vitallium Alloy; Product Number: 6260-9-528; Howmedica Osteonics Corp. Styrker Ireland; Carrigtwohill county Cork, Ireland. Made in Ireland.
- Product code
- JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
- Reason for recall
- Fracture of Stem Neck - The firm recently performed laboratory testing indicates that there may be a potential increased risk of stem neck fractures when a V40 +16mm offset femoral head is mated with certain Accolade TMZF femoral stems in patients classified as overweight or obese.
- FDA root cause
- Other
- Action
- Stryker Orthopaedics sent an Urgent Product Recall notification letters on November 9, 2007, by Federal Express with return receipt. The letter informed users of potential increased risk of stem neck fracture, when a V40 + 16mm offset femoral head is mated with certain Accolade TMZF femoral stems in patients classified as overweight or obese. The firm has asked that inventory be examined, removed and held in quarantine. A Stryker representative will contact the medical facilities to retrieve the product. If a representative has not contacted the facility to date, it is suggested that they contact Stryker at (201) 831-5972.
- Quantity in commerce
- 5,599 total units
- Distribution
- Nationwide Distribution
- Product codes and lots
- V40 COCR HEAD 28mm/+16 -- Product Number: 6260-5-528 All lot codes
FIRST ARTICLE / WEEKLY