Physio-Control Internal Defibrillator Handles. PN: 3010901-002 for use with LIFEPAK 12 and 20. PN: 805249-014...
Physio Control, Inc. began a recall of Physio-Control Internal Defibrillator Handles. PN: 3010901-002 for use with LIFEPAK 12 and 20. PN: 805249-014 thru 019 for use with LIFEPAK 9 and 9P on . The reason FDA records is "Based on testing, sterilization methods described in labeling may be ineffective and may cause damage or corrosion to the paddles or handles. If handles and/or paddles are not sterilized adequately and are used on a...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2035-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Physio Control, Inc.
- Product
- Physio-Control Internal Defibrillator Handles. PN: 3010901-002 for use with LIFEPAK 12 and 20. PN: 805249-014 thru 019 for use with LIFEPAK 9 and 9P.
- Product code
- LDD Dc-Defibrillator, Low-Energy, (Including Paddles)
- Reason for recall
- Based on testing, sterilization methods described in labeling may be ineffective and may cause damage or corrosion to the paddles or handles. If handles and/or paddles are not sterilized adequately and are used on a patient, there is a risk of infection. If handle discharge button is damaged during sterilization, defibrillation therapy may be prevented.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Starting 2/5/08, Physio-Control sent letters to customers which notifies them not to use Flash Gravity Steam, 12/88 Ethylene Oxide, STERRAD 100 machine, or Gravity Steam to sterilize. The firm recommends the use of 100% EO, STERRAD 100S machine, or Prevacuum steam for sterilization. The letter also recommends that after sterilization, the handles and paddles should be examined carefully for damage. If damaged, the product should be removed from use. The firm will follow up with a revised Sterilization Guideline document and requests posting the letter in the sterilization department in the meantime. Contact technical support at 800-442-1142 if you have questions. On 7/23/08 the firm issued new sterilization guidelines that supersede all others.
- Quantity in commerce
- 47,274 worlwide for Defibrillator Handles and Paddles
- Distribution
- Distributed to hospitals and medical centers. Nationwide, DOD, and international: ARGENTINA, BAHRAIN, BANGLADESH, BERMUDA, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, EGYPT, EL SALVADOR, FRANCE, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, INDIA, IRAN, ISRAEL, JAPAN, KOREA, LEBANON, MALAYSIA, MALTA, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, QATAR, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SRI LANKA, SYRIA, THAILAND, TURKEY, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIRGIN ISLANDS - US, YEMEN, ZIMBABWE
- Product codes and lots
- Part Number 805249 - Lot Numbers: 0002, 0003, 0004, 0006, 0007, 0008, 0009, 0011, 0101, 0102, 0103, 0104, 0105, 0110, 0112, 0201, 0202, 0203, 0205, 0206, 0208, 0209, 0211, 0304, 0305, 0306, 0307, 0308, 0401, 0402, 0404, 0405, 0412, 0502, 0503, 0506, 0507, 0509, 0603, 0604, 9508, 9509, 9608, 9609, 9610, 9702, 9706, 9708, 9709, 9807, 9809, 9811, 9812, 9902, 9906, 9907, 9908, 9909, 9912 ; Part Number 3010901 - Lot Numbers: 0003, 0004, 0005, 0006, 0007, 0008, 0009, 0012, 0101, 0102, 0103, 0104, 0106, 0107, 0108, 0109, 0110, 0111, 0112, 0201, 0202, 0204, 0205, 0206, 0207, 0208, 0209, 0210, 0211, 0212, 0301, 0302, 0303, 0304, 0305, 0306, 0307, 0308, 0310, 0312, 0401, 0402, 0403, 0404, 0405, 0406, 0407, 0411, 0412, 0501, 0502, 0503, 0504, 0505, 0509, 0510, 0511, 0512, 0602, 0603, 0604, 0605, 0606, 0608, 0610, 0611, 0612, 0701, 0702, 0704, 0705, 0710, 0711, 9808, 9809, 9811, 9812, 9902, 9904, 9905, 9907, 9909, 9910, 9911, 9912.