Z-1373-2008
Lumax Cardiac Resyncronization Therapy Defibrillator (CRT-D); Implantable, dual-chamber defibrillator; Model...
BIOTRONIK AG / initiated 2007-01-10 / Terminated / root cause: software
Biotronik Inc began a recall of Lumax Cardiac Resyncronization Therapy Defibrillator (CRT-D); Implantable, dual-chamber defibrillator; Model Number: HF-T; Catalog Number 355263; Software... on . The reason FDA records is "Timing Feature Error: An implementation error allows the usage of the V-V timing feature, which is not approved for use in the U.S". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1373-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- BIOTRONIK AG
- Product
- Lumax Cardiac Resyncronization Therapy Defibrillator (CRT-D); Implantable, dual-chamber defibrillator; Model Number: HF-T; Catalog Number 355263; Software Version 603.U; Catalog Model Number 359251; ICS 3000 units are composed of three Catalog/Model Numbers: 349527, 349530, and 355547. The model 349530 is the component which is loaded/updated with software.
- Product code
- MRM Defibrillator, Implantable, Dual-Chamber
- Reason for recall
- Timing Feature Error: An implementation error allows the usage of the V-V timing feature, which is not approved for use in the U.S.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Consignees were notified by a letter sent 1/10/07. The letter advised users to stop usage of the affected software. Users were instructed not to implant additional devices until the software program has been corrected. The recalling firm requested that all affected software be scrapped or returned to them. Users were requested update the software with a corrected version as soon as it becomes available.
- Quantity in commerce
- 53 of the 603.U discs distributed.
- Distribution
- Nationwide Distribution including states of AL, AZ, CA, CO, DE, FL, GA, IL, LA, MI, MO, NC, NJ, NV, NY, OH, OR, PA, SC, TX and WA.
- Product codes and lots
- Software version 603.U, Catalog/Model Number: 359251; ICS 300 programmers Catalog/Model Numbers: 349527, 449530 and 355547; Lumax 340 HF-T Units, Catalog Number: 355263