Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1373-2008

Lumax Cardiac Resyncronization Therapy Defibrillator (CRT-D); Implantable, dual-chamber defibrillator; Model...

BIOTRONIK AG / initiated 2007-01-10 / Terminated / root cause: software

Biotronik Inc began a recall of Lumax Cardiac Resyncronization Therapy Defibrillator (CRT-D); Implantable, dual-chamber defibrillator; Model Number: HF-T; Catalog Number 355263; Software... on . The reason FDA records is "Timing Feature Error: An implementation error allows the usage of the V-V timing feature, which is not approved for use in the U.S". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1373-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BIOTRONIK AG
Product
Lumax Cardiac Resyncronization Therapy Defibrillator (CRT-D); Implantable, dual-chamber defibrillator; Model Number: HF-T; Catalog Number 355263; Software Version 603.U; Catalog Model Number 359251; ICS 3000 units are composed of three Catalog/Model Numbers: 349527, 349530, and 355547. The model 349530 is the component which is loaded/updated with software.
Product code
MRM Defibrillator, Implantable, Dual-Chamber
Reason for recall
Timing Feature Error: An implementation error allows the usage of the V-V timing feature, which is not approved for use in the U.S.
Root cause tag
software
FDA root cause
Software design
Action
Consignees were notified by a letter sent 1/10/07. The letter advised users to stop usage of the affected software. Users were instructed not to implant additional devices until the software program has been corrected. The recalling firm requested that all affected software be scrapped or returned to them. Users were requested update the software with a corrected version as soon as it becomes available.
Quantity in commerce
53 of the 603.U discs distributed.
Distribution
Nationwide Distribution including states of AL, AZ, CA, CO, DE, FL, GA, IL, LA, MI, MO, NC, NJ, NV, NY, OH, OR, PA, SC, TX and WA.
Product codes and lots
Software version 603.U, Catalog/Model Number: 359251; ICS 300 programmers Catalog/Model Numbers: 349527, 449530 and 355547; Lumax 340 HF-T Units, Catalog Number: 355263