Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2036-2008

Internal Defibrillation Electrodes: PN: 805355-10 thru 14 and 805355-020 thru -024

Physio Control, Inc. / initiated 2008-01-29 / Terminated / root cause: labeling and UDI

Physio Control, Inc. began a recall of Internal Defibrillation Electrodes: PN: 805355-10 thru 14 and 805355-020 thru -024 on . The reason FDA records is "Based on testing, sterilization methods described in labeling may be ineffective and may cause damage or corrosion to the paddles or handles. If handles and/or paddles are not sterilized adequately and are used on a...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2036-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Physio Control, Inc.
Product
Internal Defibrillation Electrodes: PN: 805355-10 thru 14 and 805355-020 thru -024.
Product code
LDD Dc-Defibrillator, Low-Energy, (Including Paddles)
Reason for recall
Based on testing, sterilization methods described in labeling may be ineffective and may cause damage or corrosion to the paddles or handles. If handles and/or paddles are not sterilized adequately and are used on a patient, there is a risk of infection. If handle discharge button is damaged during sterilization, defibrillation therapy may be prevented.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Starting 2/5/08, Physio-Control sent letters to customers which notifies them not to use Flash Gravity Steam, 12/88 Ethylene Oxide, STERRAD 100 machine, or Gravity Steam to sterilize. The firm recommends the use of 100% EO, STERRAD 100S machine, or Prevacuum steam for sterilization. The letter also recommends that after sterilization, the handles and paddles should be examined carefully for damage. If damaged, the product should be removed from use. The firm will follow up with a revised Sterilization Guideline document and requests posting the letter in the sterilization department in the meantime. Contact technical support at 800-442-1142 if you have questions. On 7/23/08 the firm issued new sterilization guidelines that supersede all others.
Distribution
Distributed to hospitals and medical centers. Nationwide, DOD, and international: ARGENTINA, BAHRAIN, BANGLADESH, BERMUDA, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, EGYPT, EL SALVADOR, FRANCE, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, INDIA, IRAN, ISRAEL, JAPAN, KOREA, LEBANON, MALAYSIA, MALTA, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, QATAR, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SRI LANKA, SYRIA, THAILAND, TURKEY, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIRGIN ISLANDS - US, YEMEN, ZIMBABWE
Product codes and lots
Part Number 805355 - Lot Numbers: 0001, 0060, 0080, 0090, 0100, 0102, 0104, 0106, 0107, 0123, 0197, 0200, 0203, 0204, 0205, 0206, 0207, 0300, 0301, 0304, 0305, 0306, 0307, 0398, 0399, 0404, 0405, 0407, 0499, 0500, 0501, 0502, 0503, 0504, 0505, 0507, 0600, 0601, 0604, 0605, 0606, 0607, 0698, 0700, 0703, 0704, 0705, 0706, 0707, 0800, 0802, 0803, 0804, 0805, 0806, 0807, 0895, 0899, 0900, 0901, 0903, 0904, 0905, 0906, 0907, 1000, 1001, 1002, 1004, 1005, 1006, 1007, 1023, 1097, 1099, 1101, 1103, 1104, 1105, 1106, 1107, 1201, 1202, 1204, 1205, 1206, 1207, 1298, 1300, 1302, 1304, 1305, 1306, 1307, 1398, 1400, 1402, 1403, 1404, 1405, 1406, 1407, 1498, 1499, 1501, 1502, 1505, 1506, 1601, 1602, 1603, 1604, 1605, 1606, 1607, 1700, 1701, 1702, 1703, 1704, 1705, 1706, 1707, 1797, 1799, 1801, 1802, 1804, 1805, 1806, 1807, 1898, 1899, 1903, 1905, 1906, 1907, 1999, 2000, 2001, 2002, 2003, 2004, 2005, 2006, 2101, 2103, 2104, 2105, 2106, 2107, 2202, 2203, 2205, 2206, 2207, 2298, 2299, 2300, 2301, 2302, 2304, 2305, 2306, 2399, 2402, 2403, 2405, 2406, 2407, 2499, 2500, 2501, 2502, 2503, 2505, 2506, 2507, 2599, 2600, 2601, 2602, 2603, 2605, 2606, 2607, 2698, 2699, 2702, 2800, 2802, 2806, 2807, 2899, 2900, 2901, 2903, 2904, 2905, 2906, 2907, 2999, 3000, 3001, 3002, 3005, 3006, 3007, 3098, 3099, 3100, 3101, 3102, 3104, 3105, 3106, 3107, 3197, 3198, 3203, 3204, 3205, 3206, 03294, 3299, 3302, 3303, 3305, 3306, 3398, 3399, 3400, 3401, 3403, 3404, 3405, 3406, 3407, 3498, 3499, 3500, 3501, 3503, 3505, 3507, 3597, 3599, 3602, 3603, 3605, 3606, 3607, 3697, 3698, 3699, 3700, 3701, 3702, 3705, 3706, 3707, 3798, 3799, 3805, 3806, 3807, 3899, 3900, 3901, 3905, 3906, 3907, 3998, 4000, 4003, 4005, 4006, 4007, 4099, 4100, 4101, 4102, 4103, 4105, 4106, 4107, 4197, 4199, 4200, 4201, 4203, 4204, 4206, 4207, 4298, 4299, 4300, 4302, 4304, 4305, 4306, 4307, 4399, 4401, 4405, 4406, 4498, 4499, 4500, 4502, 4503, 4506, 4507, 4600, 4601, 4603, 4605, 4606, 4607, 4699, 4701, 4702, 4704, 4705, 4706, 4797, 4799, 4803, 4805, 4806, 4900, 4904, 4906, 4907, 4998, 5000, 5001, 5002, 5003, 5005, 5006, 5103, 5105, 5106, 5199, 5200, 5203, 5204, 5205, 5206, 5299, 5301.