Z-1478-2008

Allen Medical C-Flex, Polar Head Positioner Catalog Number: A-70700-A1 Product is a medical device used to...

Allen Medical Systems, Inc. / initiated 2007-11-27 / Terminated / root cause: design

Allen Medical Systems, Inc. began a recall of Allen Medical C-Flex, Polar Head Positioner Catalog Number: A-70700-A1 Product is a medical device used to maintain the position of an anesthetized patient... on . The reason FDA records is "Handle jamming in the open position, preventing the device from adequately supporting the head during spinal surgery. Possible delay in surgery, or injury". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1478-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Allen Medical Systems, Inc.
Product
Allen Medical C-Flex, Polar Head Positioner Catalog Number: A-70700-A1 Product is a medical device used to maintain the position of an anesthetized patient during surgery.
Product code
CCX Support, Patient Position
Reason for recall
Handle jamming in the open position, preventing the device from adequately supporting the head during spinal surgery. Possible delay in surgery, or injury.
Root cause tag
design
FDA root cause
Device Design
Action
Allen Medical notified accounts by telephone on 11/08/07 and the Sales Representative. The Hospitals confirmed that the device would be sent back for evaluation and upgraded.
Quantity in commerce
2
Distribution
MA, MS
Product codes and lots
Serial Numbers: A111211, A116292