IMPAX 6.2.1 Network Gateway Server, License Key. Catalog # ABC code: EXEGE000. The product is used to...
AGFA Corp. began a recall of IMPAX 6.2.1 Network Gateway Server, License Key. Catalog # ABC code: EXEGE000. The product is used to transfer medical imaging data. AGFA Corp., Greenville... on . The reason FDA records is "Failure of Data Transmission: No imaging data could be transmitted by the device, because a key value needed for data transmission was incorrectly entered in the affected units by AGFA Healthcare". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1306-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- AGFA Healthcare Corp.
- Product
- IMPAX 6.2.1 Network Gateway Server, License Key. Catalog # ABC code: EXEGE000. The product is used to transfer medical imaging data. AGFA Corp., Greenville, South Carolina 2960.
- Product code
- LMD System, Digital Image Communications, Radiological
- Reason for recall
- Failure of Data Transmission: No imaging data could be transmitted by the device, because a key value needed for data transmission was incorrectly entered in the affected units by AGFA Healthcare.
- FDA root cause
- Release of Material/Component prior to receiving test results
- Action
- On 12/11/07, AGFA Healthcare sent out an e-mail to all affected consignees informing them of the situation. The firm also arranged to install the corrected License Key on all affected units of the product.
- Quantity in commerce
- 12 units
- Distribution
- Nationwide Distribution - USA, including states of CA, GA, IA and SD.
- Product codes and lots
- ABC Code EXEGE000