Z-1751-2008
Siemens Medical Solutions, syngo Dynamics 6.0 Workplace; Picture archiving and communication system
Siemens Medical Solutions USA, Inc. / initiated 2008-02-08 / Terminated / root cause: software
Siemens Medical Solutions, USA, Inc began a recall of Siemens Medical Solutions, syngo Dynamics 6.0 Workplace; Picture archiving and communication system on . The reason FDA records is "Incorrect data: A software bug may result in a transfer of patient demographic data to a different patient's file". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1751-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Medical Solutions, syngo Dynamics 6.0 Workplace; Picture archiving and communication system
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Incorrect data: A software bug may result in a transfer of patient demographic data to a different patient's file.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens issued a customer safety advisory to users on 2/8/08 advising them of the problem and instructing them to complete and verify that the report is locked before opening and reporting on a second patient, and that a software patch will be available Spring 2008.
- Quantity in commerce
- 18
- Distribution
- Nationwide Distribution: Colorado, Florida, Georgia, Illinois, Indiana, Kansas, Michigan, Missouri, Montana, North Carolina, Texas and Utah.
- Product codes and lots
- All units with version 6.0 software.