Z-1309-2008
HOUVA III Phototherapy System with DermaSense Technology comes on one skid within 10 cardboard boxes. A...
National Biological Corp. / initiated 2008-02-19 / Terminated / root cause: software
National Biological Corp began a recall of HOUVA III Phototherapy System with DermaSense Technology comes on one skid within 10 cardboard boxes. A therapeutic product designed for individuals who... on . The reason FDA records is "Software allows operator to override "low line voltage" error warning and store light intensity value". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1309-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- National Biological Corp.
- Product
- HOUVA III Phototherapy System with DermaSense Technology comes on one skid within 10 cardboard boxes. A therapeutic product designed for individuals who require specific Ultraviolet radiation therapy for diagnosed skin disorders.
- Product code
- FTC Light, Ultraviolet, Dermatological
- Reason for recall
- Software allows operator to override "low line voltage" error warning and store light intensity value.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- National Biological sent an "Important safety reminders for your Houva DermaSense phototherapy device" letters, dated 2/11/08 to users. It states that the user should collect a "Profile" of the light energy and record it on a daily/weekly flow chart. The chart should be in plain view so each operator can clearly determine the units output. If an error message appears in the display panel, treatment should not be continued unless the phototherapy booth operator was able to identify and correct the cause of the error. Make sure both breakers are in the "ON" position before starting up the unit each morning and before giving treatment.
- Quantity in commerce
- 48 Units
- Distribution
- Worldwide distribution --- USA including states of: CA, DE, FL, GA, ID, IL, IN, MA, MI, MN, NC, NH, NY, OH, OR, PA, TX, VA, WA and WI, and country of Canada.
- Product codes and lots
- Serial #'s: 10134, 10145, 10147, 10148, 10105, 10151, 10112, 10113, 10114, 10101, 10150, 10140, 10103, 10123, 10135, 10135, 10138, 10137, 10122