High-Definition 120 MultiLeaf Collimator, Medical charged-particle radiation therapy system; Software Version...
Varian Medical Systems Oncology Systems began a recall of High-Definition 120 MultiLeaf Collimator, Medical charged-particle radiation therapy system; Software Version 7.0; Product Code: H56; Accessory to Medical... on . The reason FDA records is "Dose Calculation Error: Software anomaly may result in failure of a MLC leaf to reach planned position, potentially resulting in misadministration of dose to a patient". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1459-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- High-Definition 120 MultiLeaf Collimator, Medical charged-particle radiation therapy system; Software Version 7.0; Product Code: H56; Accessory to Medical Linear Accelerator.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Dose Calculation Error: Software anomaly may result in failure of a MLC leaf to reach planned position, potentially resulting in misadministration of dose to a patient.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Consignees were notified by a Medical Device Notification letter sent on 01/29/2008. The letter advised users to cease use of the device until a corrected version of the software is available. A representative will contact users to upgrade the application once the revised software has been validated and released for distribution.
- Quantity in commerce
- 3 units subject to correction
- Distribution
- Nationwide Distribution including states of NC, FL and CA.
- Product codes and lots
- Serial Number 1009.