GE Healthcare Proteus XR/a Radiographic X-ray System, Model Number 2259976; previously marketed as the...
GE Healthcare began a recall of GE Healthcare Proteus XR/a Radiographic X-ray System, Model Number 2259976; previously marketed as the Silhouette FC, GE Healthcare, Waukesha, WI 53188 on . The reason FDA records is "The warning label that is required by 21 CFR 1020.30(j) was not on the control console of the Proteus XR/a Radiographic X-ray System, Model 2259976". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0842-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Proteus XR/a Radiographic X-ray System, Model Number 2259976; previously marketed as the Silhouette FC, GE Healthcare, Waukesha, WI 53188.
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- The warning label that is required by 21 CFR 1020.30(j) was not on the control console of the Proteus XR/a Radiographic X-ray System, Model 2259976.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- GE issued a Field Modification Instruction to their field service representatives, released in December 2007, with instructions to replace the labels on the console with the appropriate labeling. Consignees will be visited by GE field service representatives beginning December 2007. All consignees will have the required labels affixed to their units by July 31, 2008.
- Quantity in commerce
- 400 Units
- Distribution
- Worldwide Distribution, including USA, Canada, Mexico, France, Hong Kong, Italy, Korea, China, Poland, UK, New Zealand, Saudi Arabia, Taiwan, Turkey, and Australia.
- Product codes and lots
- 00000010838HL1, 00000011061HL9, 00000011067HL6, 00000011185HL6, 00000012500HL5, 00000012667HL2, 00000013129HL2, 00000013372HL8, 00000013385HL0, 00000013543HL4, 00000013557HL4, 00000013800HL8, 00000014099HL6, 00000014104HL4, 00000014319HL8, 00000014327HL1, 00000014667HL0, 00000014677HL9, 00000015076HL3, 00000015383HL3, 00000015386HL6, 00000015400HL5, 00000015738HL8, 00000015744HL6, 00000015745HL3, 00000015750HL3, 00000015760HL2, 00000016174HL5, 00000016383HL2, 00000016386HL5, 00000016696HL7, 00000016707HL2, 00000016708HL0, 00000016709HL8, 00000016930HL0, 00000016954HL0, 00000016955HL7, 00000017119HL9, 00000017130HL6, 00000017131HL4, 00000017476HL3, 00000017477HL1, 00000017478HL9, 00000017480HL5, 00000017490HL4, 00000017492HL0, 00000017493HL8, 00000017599HL2, 00000017604HL0, 00000017608HL1, 00000017610HL7, 00000017613HL1, 00000017616HL4, 00000017788HL1, 00000017801HL2, 00000018476HL2, 00000018479HL6, 00000018483HL8, 00000019041HL3, 00000019048HL8, 00000019095HL9, 00000019098HL3, 00000019104HL9, 00000019110HL6, 00000019241HL9, 00000019244HL3, 00000019413HL4, 00000019419HL1, 00000019489HL4, 00000019496HL9, 00000019499HL3, 00000019501HL6, 00000019502HL4, 00000019669HL1, 00000019801HL0, 00000019886HL1, 00000019894HL5, 00000019959HL6, 00000020091HL5, 00000020092HL3, 00000020239HL0, 00000020243HL2, 00000020374HL5, 00000020967HL6, 00000021064HL1, 00000021197HL9, 00000021201HL9, 00000021204HL3, 00000021343HL9, 00000021344HL7, 00000021349HL6, 00000021552HL5, 00000021911HL3, 00000021921HL2, 00000021938HL6, 00000021953HL5, 00000022033HL5, 00000022333HL9, 00000022337HL0, 00000022412HL1, 00000022477HL4, 00000022600HL1, 00000022795HL9, 00000022931HL0, 00000023116HL7, 00000023122HL5, 00000023126HL6, 00000023412HL0, 00000023415HL3, 00000024386HL5, 00000024483HL0, 00000024588HL6, 00000024939HL1, 00000024940HL9, 00000025426HL8, 00000025475HL5, 00000025476HL3, 00000025477HL1, 00000025572HL9, 00000027104HL9, 00000027867HL1, 00000028057HL8, 00000028453HL9, 00000028524HL7, 00000028845HL6, 00000029098HL1, 00000029145HL0, 00000029502HL2, 00000029550HL1, 00000029634HL3, 00000029640HL0, 00000029969HL3, 00000030108HL5, 00000030113HL5, 00000030229HL9, 00000030232HL3, 00000030236HL4, 00000030633HL2, 00000030848HL6, 00000030920HL3, 00000030921HL1, 00000030976HL5, 00000031172HL0, 00000031175HL3, 00000031565HL5, 00000031702HL4, 00000031811HL3, 00000031928HL5, 00000031929HL3, 00000031930HL1, 00000032309HL7, 00000032311HL3, 00000032317HL0, 00000032319HL6, 00000032322HL0, 00000032850HL0, 00000032912HL8, 00000032916HL9, 00000032919HL3, 00000032920HL1, 00000033330HL2, 00000033388HL0, 00000033391HL4, 00000033408HL6, 00000033437HL5, 00000033509HL1, 00000033679HL2, 00000033681HL8, 00000033682HL6, 00000033727HL9, 00000033731HL1, 00000033884HL8, 00000033970HL5, 00000033971HL3, 00000033983HL8, 00000033984HL6, 00000034218HL8, 00000034420HL0, 00000034469HL7, 00000034470HL5, 00000034520HL7, 00000034644HL5, 00000034694HL0, 00000034762HL5, 00000034764HL1, 00000034913HL4, 00000034938HL1, 00000035052HL0, 00000035101HL5, 00000035159HL3, 00000035191HL6, 00000035244HL3, 00000035246HL8, 00000035247HL6, 00000035329HL2, 00000035445HL6, 00000035544HL6, 00000035547HL9, 00000035632HL9, 00000035641HL0, 00000036038HL8, 00000036068HL5, 00000036071HL9, 00000036100HL6, 00000036126HL1, 00000036128HL7, 00000036265HL7, 00000036362HL2, 00000036363HL0, 00000036431HL5, 00000036624HL5, 00000036625HL2, 00000036674HL0, 00000036698HL9, 00000036702HL9, 00000036882HL9, 00000036983HL5, 00000037104HL7, 00000037123HL7, 00000037351HL4, 00000037352HL2, 00000037353HL0, 00000037396HL9, 00000037596HL4, 00000037919HL8, 00000038423HL0, 00000038771HL2, 00000038772HL0, 00000039125HL0, 00000039581HL4, 00000039612HL7, 00000040366HL7, 00000040493HL9, 00000040728HL8, 00000040865HL8, 00000041290HL8, 00000041291HL6, 00000041511HL7, 00000041768HL3, 00000042367HL3, 00000042649HL4, 00000042650HL2, 00000042651HL0, 00000042701HL3, 00000042880HL5, 00000043029HL8, 00000043184HL1, 00000043265HL8, 00000043665HL9, 00000043727HL7.00000043747HL5, 00000043751HL7, 00000044548HL6, 00000045475HL1, 00000045651HL7, 00000015925HL1, 00000019969HL5, 00000024476HL4, 00000027957HL0, 00000030025HL1, 00000030028HL5, 00000032711HL4, 00000035548HL7, 00000036428HL1, 00000036883HL7, 00000020371HL1, 00000021582HL2, 00000031922HL8, 00000033316HL1, 00000042706HL2, 00000020735HL7, 00000022513HL6, 00000022926HL0, 00000024376HL6, 00000031909HL5, 00000040521HL7, 00000041017HL5, 00000011544HL4, 00000014336HL2, 00000016171HL1, 00000016545HL6, 00000017123HL1, 00000017483HL9, 00000023189HL4, 00000028833HL2, 00000030875HL9, 00000030924HL5, 00000031916HL0, 00000031918HL6, 00000032909HL4, 00000033730HL3, 00000033974HL7, 00000034268HL3, 00000036165HL9, 00000042001HL8, 00000016177HL8, 00000031554HL9, 00000036626HL0, 00000013383HL5, 00000014323HL0, 00000019485HL2, 00000020232HL5, 00000010836HL5, 00000019679HL0, 00000023546HL5, 00000023547HL3, 00000030926HL0, 00000041996HL0, 00000042388HL9, 00000042537HL1, 00000011063HL5, 00000015391HL6, 00000017138HL9, 00000021058HL3, 00000030918HL7, 00000040086HL1, 00000010328HL3, 00000013375HL1, 00000014335HL4, 00000015065HL6, 00000015081HL3, 00000015747HL9, 00000016371HL7, 00000016701HL5, 00000017125HL6, 00000017481HL3, 00000019482HL9, 00000019486HL0, 00000019678HL2, 00000019680HL8, 00000019788HL9, 00000020081HL6, 00000020224HL2, 00000021639HL0, 00000022032HL7, 00000023118HL3, 00000023181HL1, 00000024432HL7, 00000029259HL9, 00000031908HL7, 00000031914HL5, 00000032231HL3, 00000032897HL1, 00000033724HL6, 00000033725HL3, 00000033968HL9, 00000034260HL0, 00000034459HL8, 00000034462HL2, 00000034783HL1, 00000035377HL1, 00000035378HL9, 00000036563HL5, 00000037254HL0, 00000038190HL5, 00000016175HL2, 00000019042HL1, 00000013796HL8, 00000011182HL3, 00000032313HL9, 00000012475HL0, 00000020222HL6, 00000029743HL2, 00000033726HL1, 00000020732HL4, 00000020737HL3, 00000012674HL8, 00000021551HL7, 00000031689HL3, 00000010651HL8, 00000015757HL8, 00000022520HL1, 00000015082HL1, 00000015153HL0, 00000021644HL0, 00000013373HL6, 00000015158HL9, 00000022606HL8, 00000022607HL6, 00000024489HL2, 00000024590HL2, 00000024671HL0, 00000024689HL2, 00000030446HL9, 00000030460HL0, 00000030927HL8, 00000036429HL9, 00000015387HL4, 00000039893HL3, 00000039895HL8, 00000039962HL6, 00000040002HL8, 00000040072HL1, 00000040164HL6, 00000040167HL9, 00000022402HL2, 00000036266HL5, 00000036914HL0, 00000037252HL4, 00000013564HL0.