Z-1310-2008

Cardiovascular Procedure Kit, P/N: 70932-01, HCSG-St. Mary's Hospital, Richmond, VA , X-coated Pump and Table...

Terumo Cardiovascular Systems Corporation / initiated 2008-01-29 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corp began a recall of Cardiovascular Procedure Kit, P/N: 70932-01, HCSG-St. Mary's Hospital, Richmond, VA , X-coated Pump and Table Pack, Terumo Cardiovascular, Ashland, MA (the... on . The reason FDA records is "Loose Flash at tip of Weighted Flexible Sucker: A loose strand of stainless steel left on the device during manufacturing by an outside supplier may result in an unretrieved metal fragment remaining in the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1310-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Cardiovascular Procedure Kit, P/N: 70932-01, HCSG-St. Mary's Hospital, Richmond, VA , X-coated Pump and Table Pack, Terumo Cardiovascular, Ashland, MA (the X-coated pump and table pack contains a weighted flexible sucker, Catalog No.: 804113)
Product code
OEZ Cardiovascular Procedure Kit
Reason for recall
Loose Flash at tip of Weighted Flexible Sucker: A loose strand of stainless steel left on the device during manufacturing by an outside supplier may result in an unretrieved metal fragment remaining in the patient.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Consignees were notified via recall letter, Urgent Medical Device Removal, dated 1/29/08 via fax or email to cease using the weighted sucker in the packs and to use an alternative sucker until replacement product arrives from Terumo.
Quantity in commerce
739
Distribution
Nationwide-USA including states of Missouri, Tennessee and Wisconsin-Custom Kits for 3 hospitals.
Product codes and lots
Lot Numbers: GC06, GC27, GD27, GG19, GK07, GL05, GM02, GM30, HA02, HC05, HD12, HE23, HF29, HH23, HK27, HM22, HP03 and KA14.