Z-1182-2008

System 83 Plus Washer-Disinfector, Endoscope use, DOS based system, Custom Ultrasonics Inc., Buckingham, PA

Wassenburg Medical B.V. / initiated 2008-02-11 / Terminated / root cause: labeling and UDI

Custom Ultrasonics, Inc. began a recall of System 83 Plus Washer-Disinfector, Endoscope use, DOS based system, Custom Ultrasonics Inc., Buckingham, PA on . The reason FDA records is "Inadvertent selection of the "wash" cycle". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1182-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Wassenburg Medical B.V.
Product
System 83 Plus Washer-Disinfector, Endoscope use, DOS based system, Custom Ultrasonics Inc., Buckingham, PA
Product code
KOG
Reason for recall
Inadvertent selection of the "wash" cycle
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
The recalling firm issued a Product Correction letter dated 2/11/08 to its customers informing them of the problem and the correction choices. The letter states that a key cover is available to reduce the error in the future and will be provided to each users upon return of the recall letter notification questionnaire.
Quantity in commerce
3,190 units
Distribution
Worldwide; USA medical facilities and physicians and Canada
Product codes and lots
All Serial Numbers