Z-1383-2008

Warfarin Metabolism Nucleic Acid Test Cartridge; Part Number: 20-006-002; an in vitro diagnostic drug...

Luminex Corporation / initiated 2008-02-25 / Terminated

Nanosphere, Inc. began a recall of Warfarin Metabolism Nucleic Acid Test Cartridge; Part Number: 20-006-002; an in vitro diagnostic drug metabolizing enzyme genotyping system; Nanosphere, Inc... on . The reason FDA records is "Aberrant Results: A warfarin 2C9*2 mutant capture signal was found to be high enough to result in an aberrant result (i.e., wild-type called heterozygous)". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1383-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Luminex Corporation
Product
Warfarin Metabolism Nucleic Acid Test Cartridge; Part Number: 20-006-002; an in vitro diagnostic drug metabolizing enzyme genotyping system; Nanosphere, Inc., 4088 Commercial Ave., Northbrook, IL 60062
Product code
ODW Cytochrome P450 2c9 (Cyp450 2c9) Drug Metabolizing Enzyme Genotyping System
Reason for recall
Aberrant Results: A warfarin 2C9*2 mutant capture signal was found to be high enough to result in an aberrant result (i.e., wild-type called heterozygous).
FDA root cause
Other
Action
Nanosphere sent Urgent Device Recall letters dated 2/25/08 to the direct accounts on the same date, informing them of the potential for aberrant results with the use of the affected lots of cartridges. The accounts were requested to examine their stocks for the affected lots of cartridges, segregate those found and return them to Nanosphere for replacement. Any questions can be directed to Dr. Gregg Shipp at 847-400-9115.
Quantity in commerce
624 units
Distribution
Nationwide; USA including states of Kentucky, Michigan, Tennessee, Texas, Louisiana, Connecticut, Illinois, and California.
Product codes and lots
Lot Numbers 121907002B, 010708002A, 012108002A, 012108002C, 012208002C, 012408002B, 012808002A, and 020608002A