Z-1383-2008
Warfarin Metabolism Nucleic Acid Test Cartridge; Part Number: 20-006-002; an in vitro diagnostic drug...
Luminex Corporation / initiated 2008-02-25 / Terminated
Nanosphere, Inc. began a recall of Warfarin Metabolism Nucleic Acid Test Cartridge; Part Number: 20-006-002; an in vitro diagnostic drug metabolizing enzyme genotyping system; Nanosphere, Inc... on . The reason FDA records is "Aberrant Results: A warfarin 2C9*2 mutant capture signal was found to be high enough to result in an aberrant result (i.e., wild-type called heterozygous)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1383-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Luminex Corporation
- Product
- Warfarin Metabolism Nucleic Acid Test Cartridge; Part Number: 20-006-002; an in vitro diagnostic drug metabolizing enzyme genotyping system; Nanosphere, Inc., 4088 Commercial Ave., Northbrook, IL 60062
- Product code
- ODW Cytochrome P450 2c9 (Cyp450 2c9) Drug Metabolizing Enzyme Genotyping System
- Reason for recall
- Aberrant Results: A warfarin 2C9*2 mutant capture signal was found to be high enough to result in an aberrant result (i.e., wild-type called heterozygous).
- FDA root cause
- Other
- Action
- Nanosphere sent Urgent Device Recall letters dated 2/25/08 to the direct accounts on the same date, informing them of the potential for aberrant results with the use of the affected lots of cartridges. The accounts were requested to examine their stocks for the affected lots of cartridges, segregate those found and return them to Nanosphere for replacement. Any questions can be directed to Dr. Gregg Shipp at 847-400-9115.
- Quantity in commerce
- 624 units
- Distribution
- Nationwide; USA including states of Kentucky, Michigan, Tennessee, Texas, Louisiana, Connecticut, Illinois, and California.
- Product codes and lots
- Lot Numbers 121907002B, 010708002A, 012108002A, 012108002C, 012208002C, 012408002B, 012808002A, and 020608002A