Da Vinci S 8mm Instrument Cannula with outlet for use with the Da Vinci Surgical System; Model IS2000, Part...
Intuitive Surgical, Inc. began a recall of Da Vinci S 8mm Instrument Cannula with outlet for use with the Da Vinci Surgical System; Model IS2000, Part Numbers: 400254-01, 400255-01; Used in endoscopic... on . The reason FDA records is "(1) Incorrect dimension on Luer on smoke evacuation cannulae not allowing for secure attachment function. (2) Incorrect labeling. External labeling lot number differs from lot number etched on the cannula". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1348-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- Da Vinci S 8mm Instrument Cannula with outlet for use with the Da Vinci Surgical System; Model IS2000, Part Numbers: 400254-01, 400255-01; Used in endoscopic surgery to establish a port of entry for Intuitive Surgical's EndoWrist Instruments. Product is distributed by Intuitive Surgical, Inc., Sunnyvale, CA
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- (1) Incorrect dimension on Luer on smoke evacuation cannulae not allowing for secure attachment function. (2) Incorrect labeling. External labeling lot number differs from lot number etched on the cannula.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 1/4/08, the firm initiated the recall and its notification was via Urgent Device Recall letters explaining the reason for the recall, requesting immediate discontinued use, segregate non-conforming product in a secure area, and a Customer Service Rep will make contact in order to arrange for the retrieval of the material.
- Quantity in commerce
- 39 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 13730-01-70807, 13793-71106, 13730-02-70807, WI073202, WI074501, WI073201