Z-1471-2008
Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System, Catalog number AS4010B
Biogenex Laboratories, Inc. / initiated 2008-01-24 / Terminated / root cause: design
Biogenex Laboratories began a recall of Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System, Catalog number AS4010B on . The reason FDA records is "Bulk reagent dispensing mechanism may break resulting in reagent spill and possible exposure to chemicals". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1471-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Biogenex Laboratories, Inc.
- Product
- Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System, Catalog number AS4010B
- Product code
- KEY Stainer, Tissue, Automated
- Reason for recall
- Bulk reagent dispensing mechanism may break resulting in reagent spill and possible exposure to chemicals.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- The firm initiated written notification on 01/24/2008. The firm has asked its consignee to inform its consignees of the recall.
- Quantity in commerce
- 10 units distributed.
- Distribution
- Product was distributed to a single distributor, which further distributed the product.
- Product codes and lots
- Serial numbers AS40211, AS40226, AS40230, AS40231, AS40232, AS40233, AS40234, AS40235, AS40236, AS40237. These serial numbers encompass ALL products distributed.