Z-1471-2008

Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System, Catalog number AS4010B

Biogenex Laboratories, Inc. / initiated 2008-01-24 / Terminated / root cause: design

Biogenex Laboratories began a recall of Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System, Catalog number AS4010B on . The reason FDA records is "Bulk reagent dispensing mechanism may break resulting in reagent spill and possible exposure to chemicals". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1471-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Biogenex Laboratories, Inc.
Product
Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System, Catalog number AS4010B
Product code
KEY Stainer, Tissue, Automated
Reason for recall
Bulk reagent dispensing mechanism may break resulting in reagent spill and possible exposure to chemicals.
Root cause tag
design
FDA root cause
Component design/selection
Action
The firm initiated written notification on 01/24/2008. The firm has asked its consignee to inform its consignees of the recall.
Quantity in commerce
10 units distributed.
Distribution
Product was distributed to a single distributor, which further distributed the product.
Product codes and lots
Serial numbers AS40211, AS40226, AS40230, AS40231, AS40232, AS40233, AS40234, AS40235, AS40236, AS40237. These serial numbers encompass ALL products distributed.