EvoTech Endoscope Cleaner and Reprocessor, Product Code 50004-002 Indicated for use with high level...
Advanced Sterilization Products began a recall of EvoTech Endoscope Cleaner and Reprocessor, Product Code 50004-002 Indicated for use with high level disinfectant CIDEX OPA Concentrate and an enzymatic... on . The reason FDA records is "Cleaning cycles are being cancelled. Specific issues are: a) Water heater tank is rusting resulting in the in-line 0.2 micron filter prematurely clogging, b) Interface board malfunction resulting in a premature lid...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1369-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Advanced Sterilization Products
- Product
- EvoTech Endoscope Cleaner and Reprocessor, Product Code 50004-002 Indicated for use with high level disinfectant CIDEX OPA Concentrate and an enzymatic detergent, CIDEZYME GI, to achieve cleaning and high level disinfection of heat sensitive semi-critical endoscopes.
- Product code
- FEB Accessories, Cleaning, For Endoscope
- Reason for recall
- Cleaning cycles are being cancelled. Specific issues are: a) Water heater tank is rusting resulting in the in-line 0.2 micron filter prematurely clogging, b) Interface board malfunction resulting in a premature lid opening, and c) Minor fluid leaks within the system (commonly at the 0.2 micron filter attachment points).
- FDA root cause
- Other
- Action
- Feburary 13, 2008 ASP issued an Urgent Product Recall Letter. ASP recommends that customers use an alternative reprocessing method until they return the system. However, if they do not have an alternate method for processing endoscopes and choose to continue using the EvoTech ECR System during this time, endoscopes processed through completed cycles in accordance with the User's Guide will be appropriately cleaned and highlevel disinfected. ASP is prepared to tailor an alternative solution to minimize the interruption to the customer's processing needs.
- Quantity in commerce
- 2 units
- Distribution
- Worldwide Distribution --- including USA and countries of Canada and Japan.
- Product codes and lots
- Product Code 50004-002; Serial Numbers 50004002070001 and 50004002070004