Z-1472-2008
Varta Easypack XL Lithium Polymer Batteries (P/N 41772 and 42215) Used with: Nova Biomedical StatStrip...
Nova Biomedical Corporation / initiated 2008-02-21 / Terminated / root cause: design
Nova Biomedical Corporation began a recall of Varta Easypack XL Lithium Polymer Batteries (P/N 41772 and 42215) Used with: Nova Biomedical StatStrip Glucose Hospital Meter (P/N 42224) and Nova Biomedical... on . The reason FDA records is "Lithium batteries may fail if the glucose or creatinine meters have been dropped". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1472-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nova Biomedical Corporation
- Product
- Varta Easypack XL Lithium Polymer Batteries (P/N 41772 and 42215) Used with: Nova Biomedical StatStrip Glucose Hospital Meter (P/N 42224) and Nova Biomedical StatSensor Creatinine Hospital Meter and EZ CHEM Creatinine Meter (P/N 9600). Statstrip Meter an In vitro diagnostic for use by health care professionals in the quantitative determination of glucose (GLU) in whole blood. EZ Chem In vitro diagnostic for the quantitative measurement of creatinine in capillary, venous or arterial blood.
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- Lithium batteries may fail if the glucose or creatinine meters have been dropped.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Nova Biomedical notified consignees by telephone and followed by the Recall Notification letter for the Vata Easy Pack XL Lithium Polymer Battery, dated 2/2108 and mailed 2/26/08. Users are notified of the problem with the battery in the Nova Stat Strip Blood Glucose Hospital Meter System and is replacing batteries by on site visit. E-Z-EM , Inc. notified their customers by letter dated 3/17/08. Contact Nova Biomedical Corporation at 1-800-545-6682 for assistance.
- Quantity in commerce
- 4432 batteries
- Distribution
- Worldwide Distribution --- including USA and countries of Canada, Taiwan, Hungary, Malaysia, Australia, Russia, Thailand, New Zealand, Austria, France, Germany, UK, China, South Africa, Croatian, Peru, and Czech Republic.
- Product codes and lots
- Batteries coded: XX07 and earlier The Nova Biomedical products that contain the lithium batteries include the following part numbers: Part # 41772 - Battery Assy, Part # 42215 - Battery Replacement pack, Part # 42224 - Kit Meter, Part # 43735 - Kit Meter EZ CHEM, Part # 43773 - Kit Meter, Part # 43811 - Battery Replacement Pack EZ CHEM, Part # 44013 - Kit Meter, Part # 44014 - Kit Meter and Part # 44251 - Kit Meter.