Stryker Digital Capture System Ultra (SDC Ultra), Model Number 240-050-988, Stryker Endoscopy, image and...
Stryker Endoscopy began a recall of Stryker Digital Capture System Ultra (SDC Ultra), Model Number 240-050-988, Stryker Endoscopy, image and video archiving system on . The reason FDA records is "Patient Identification numbers and names may be applied to images from another patient due to a software issue. This failure mode can only occur when the pre-load function of the SDC ultra is used". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1316-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker Digital Capture System Ultra (SDC Ultra), Model Number 240-050-988, Stryker Endoscopy, image and video archiving system.
- Product code
- LMB Device, Digital Image Storage, Radiological
- Reason for recall
- Patient Identification numbers and names may be applied to images from another patient due to a software issue. This failure mode can only occur when the pre-load function of the SDC ultra is used.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Stryker Endoscopy mailed Urgent Device Correction notices, dated February 14, 2008, to all consignees, and sales representatives will be supplied with a fix to distribute to the user base.
- Quantity in commerce
- 814 units distributed
- Distribution
- Worldwide distribution: USA, Canada, Greece, and Germany.
- Product codes and lots
- Serial numbers 07H041654 - 08A069464, PP1.04