Z-1532-2008
Quickie Groove Power Wheelchair with QC (non-expandable) electronics, Model number EIPW13
Sunrise Medical, Inc. / initiated 2008-01-11 / Terminated / root cause: software
Sunrise Medical Inc began a recall of Quickie Groove Power Wheelchair with QC (non-expandable) electronics, Model number EIPW13 on . The reason FDA records is "Product experiences a longer than expected stopping distance, which may cause injury to the user, objects, or the chair itself". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1532-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sunrise Medical, Inc.
- Product
- Quickie Groove Power Wheelchair with QC (non-expandable) electronics, Model number EIPW13.
- Product code
- ITI Wheelchair, Powered
- Reason for recall
- Product experiences a longer than expected stopping distance, which may cause injury to the user, objects, or the chair itself.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Initial phone contact began on January 11, 2008. Medical device correction letters were also sent to affected dealers.
- Quantity in commerce
- 141 units total
- Distribution
- Product was distributed to 126 retailers nationwide throughout the USA, as well as Canada and Australia.
- Product codes and lots
- All wheelchairs with the QC Motor controller units with version 2.0 parameter or *.SUN file software.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.