Z-1892-2008

EKOS EndoWave Infusion System consisting of a Drug Delivery Catheter and an Ultrasound Core. Labeled as...

EKOS Corporation / initiated 2008-02-20 / Terminated / root cause: manufacturing process

EKOS Corporation began a recall of EKOS EndoWave Infusion System consisting of a Drug Delivery Catheter and an Ultrasound Core. Labeled as Sterile, Single Use Only. The device is intended for... on . The reason FDA records is "Product mislabeled - 6 cm treatment zone catheter had been built using 50 cm treatment zone drilled tubing. Consequently, drug was delivered outside of the treatment zone". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1892-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
EKOS Corporation
Product
EKOS EndoWave Infusion System consisting of a Drug Delivery Catheter and an Ultrasound Core. Labeled as Sterile, Single Use Only. The device is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.
Product code
KRA Catheter, Continuous Flush
Reason for recall
Product mislabeled - 6 cm treatment zone catheter had been built using 50 cm treatment zone drilled tubing. Consequently, drug was delivered outside of the treatment zone.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 02/20/08 firm notified customers by phone or email informing them of the issue and notifying them to segregate affected product and hold for return. Letters were sent via FedEx or email, or were hand delivered by firm representative to customers on 02/21/08 informing them of the issue and requesting return of affected product for replacement. Contact Jocelyn Kersten at 425-415-3132 if you have questions.
Quantity in commerce
10 units
Distribution
Product was distributed to hospitals in the following states: WA, NC, IL, OH, NY, MI, and KY with no foreign distribution.
Product codes and lots
071126006-001, 071126006-002, 071126006-003, 071126006-005, 071204008-001, 071128015-001, 071128015-002, 071128015-003, 071128015-004, 071128015-005, Item number 500-52106, expiration date 2009-10