Acuity with Conebeam Computed Tomography; Radiation therapy simulation system; Model Number: H77
Varian Medical Systems Oncology Systems began a recall of Acuity with Conebeam Computed Tomography; Radiation therapy simulation system; Model Number: H77 on . The reason FDA records is "Incorrect Image Orientation: The image orientation tag may not be set correctly if the patient has been scanned with an orientation other than Head First Spine. When such images are imported to a treatment planning...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1460-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Acuity with Conebeam Computed Tomography; Radiation therapy simulation system; Model Number: H77.
- Product code
- KPQ System, Simulation, Radiation Therapy
- Reason for recall
- Incorrect Image Orientation: The image orientation tag may not be set correctly if the patient has been scanned with an orientation other than Head First Spine. When such images are imported to a treatment planning system, they may appear with either a mirrored or a rotated view.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Consignees were notified by an Urgent Medical Device Correction letter issued on 02/05/2008. The letter provided recommended actions for users to follow in order to avoid the software error. The letter informed users that a service pack is being created to correct the issue and that the recalling firm will notify users when it becomes available. For additional information, contact 888-827-4265.
- Quantity in commerce
- 16 UNITS
- Distribution
- Worldwide Distribution--USA including states of CA, MA and PA, and countries of UK, Chile, Belgium, Canada and Germany.
- Product codes and lots
- Serial Numbers: H770002-H780001, H770012-H780058, H770030-H780034, H770055-H780046, H770068-H780010, H770107-H780050, H770113-H780060, H770119-H780035, H770134-H780026, H770143-H780039, H770159-H780062, H770170-H780056, H770184-H780061, H770212-H780074, H770245-H780086 and H770279-H780105.