Z-1515-2008
SEQUENCER, MOSAIQ versions 1.3 and 1.4 with Elekta and Siemens Machines, Model number 40003-001000-RO...
Elekta Limited / initiated 2008-02-14 / Terminated / root cause: software
Impac Medical Systems Inc began a recall of SEQUENCER, MOSAIQ versions 1.3 and 1.4 with Elekta and Siemens Machines, Model number 40003-001000-RO, Manufactured by IMPAC Medical Systems, Sunnyvale, CA... on . The reason FDA records is "Software issue may result in change to intended treatment field, potentially resulting in mistreatment". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1515-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- SEQUENCER, MOSAIQ versions 1.3 and 1.4 with Elekta and Siemens Machines, Model number 40003-001000-RO, Manufactured by IMPAC Medical Systems, Sunnyvale, CA. Used to facilitate the delivery of defined radiotherapy treatment plans.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Software issue may result in change to intended treatment field, potentially resulting in mistreatment.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Firm issued field advisory to its consignees on 02/14/2008. Phone calls will also be made to affected customers to ensure reception and understanding of the field notification. Firm advises users with affected product to review the jaw positions of IMRT fields to ensure that they match the RT Plan after an Apply Limits or Machine Change occurs, until software at their site can be updated to the corrected version. A software patch is under construction.
- Quantity in commerce
- 36 units
- Distribution
- Worldwide Distribution --- USA including states of OH, OK, NJ, MN, OR, TN, NY, CA, AZ, NV, IA, GA, AL, MD, MI, PA, IN, WA, FL, TX, CO, and countries to Canada, Australia, Germany, Greece and Sweden.
- Product codes and lots
- MOSAIQ versions 1.30 and 1.40 with Elekta iCom or Siemens DMIP