Z-1541-2008
IntelliVue Clinical Information Portfolio Critical Care Release D, Version D.00.00 Product Number: 865047...
Philips North America LLC / initiated 2008-03-04 / Terminated / root cause: software
Philips Medical Systems began a recall of IntelliVue Clinical Information Portfolio Critical Care Release D, Version D.00.00 Product Number: 865047. This device is a clinical information system... on . The reason FDA records is "Medications prescribed for one patient were printed on the record of another patient". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1541-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- IntelliVue Clinical Information Portfolio Critical Care Release D, Version D.00.00 Product Number: 865047. This device is a clinical information system intended for use in data collection, storage, and management with independent bedside devices, and other systems that are connected either directly or through networks. It is used when there is a need for generating patient records and computation of drug dosage.
- Product code
- DXJ Display, Cathode-Ray Tube, Medical
- Reason for recall
- Medications prescribed for one patient were printed on the record of another patient.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips Medical issued letter notification, dated 29 Feb 2008 to customers titled; "Urgent Medical Device Correction Notice". Customers were asked to not solely rely on the Worklist to remind them when to administer orders or interventions until the software has been upgraded. Customers should review the reminders and verify orders and interventions on the Medication Administration Record and Administration record. The correction will be in the form of a software upgrade and will be performed remotely by Philips service personnel.
- Quantity in commerce
- 138 for products
- Distribution
- Nationwide Canada, Australia, Austria, Belgium, Denmark., France, Finland, Germany, Great Britain, Italy, Malaysia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, and UAE
- Product codes and lots
- Version D.00.00