3DKNEE Tibial Insert Trial, Size 2, 13mm Right; Part #801-01-640; manufactured by Encore Medical, L.P...
Encore Medical, Lp began a recall of 3DKNEE Tibial Insert Trial, Size 2, 13mm Right; Part #801-01-640; manufactured by Encore Medical, L.P., Austin, TX 78758. Used in surgery to determine the... on . The reason FDA records is "Tibial insert trial incorrectly marked for thickness; device thickness marked correctly as 13mm on the underside of the trial, but is incorrectly marked as 15mm on the side". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1539-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- 3DKNEE Tibial Insert Trial, Size 2, 13mm Right; Part #801-01-640; manufactured by Encore Medical, L.P., Austin, TX 78758. Used in surgery to determine the correct size of 3DKnee component to implant that would allow for the best range of motion and joint stability.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Tibial insert trial incorrectly marked for thickness; device thickness marked correctly as 13mm on the underside of the trial, but is incorrectly marked as 15mm on the side.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Firm notified consignee by phone and asked them to return any outstanding product on 02/20/08.
- Quantity in commerce
- 6 units.
- Distribution
- Distribution --- country of Germany.
- Product codes and lots
- Lot #53906407