ArthroCare brand Topaz XL ArthroWand with Integrated Cable Wand, Catalog Number: AC4045-01; Product is...
ArthroCare Corporation began a recall of ArthroCare brand Topaz XL ArthroWand with Integrated Cable Wand, Catalog Number: AC4045-01; Product is manufactured and distributed by ArthroCare Corporation... on . The reason FDA records is "Devices for which sterility may be compromised as evidenced by punctures in the plastic tray during shipping. The risk to the patient is a potential infection subsequent to the surgical procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1657-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ArthroCare Corporation
- Product
- ArthroCare brand Topaz XL ArthroWand with Integrated Cable Wand, Catalog Number: AC4045-01; Product is manufactured and distributed by ArthroCare Corporation, Sunnyvale, CA - Sterile Surgical Device. Product is indicated for ablation and debridement of tendons in orthopedic procedures.
- Product code
- HRX Arthroscope
- Reason for recall
- Devices for which sterility may be compromised as evidenced by punctures in the plastic tray during shipping. The risk to the patient is a potential infection subsequent to the surgical procedure.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Urgent Product Recall letters (Dated January 31, 2008) were issued to end users of the device (medical professionals) and direct distributors. The letter instructs users to return any affected products.
- Quantity in commerce
- 2,129
- Distribution
- Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, NC, ND, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and countries of Sweden, Australia, Germany, Italy, UK, Austria, Saudi Arabia, Belgium, Portugal, UAE, Kuwait, Singapore, Ireland, Spain, Finland, South Korea, Russia and Vietnam.
- Product codes and lots
- Lot Numbers: A603250-A, A609450-A, A610250-A, A615750-A, A616850-A, A627860-A, A627860-B, A604670-A, A615870-A, A620170-A, A623270-A, and A634670-A.