Z-1627-2008

Gammas Set Screwdriver Instrument Set, Basic Stryker Trauma GmbH, Germany; Distributed by Howmedica Osteonics...

Howmedica Osteonics Corp / initiated 2008-02-22 / Terminated

Stryker Howmedica Osteonics Corp. began a recall of Gammas Set Screwdriver Instrument Set, Basic Stryker Trauma GmbH, Germany; Distributed by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 -... on . The reason FDA records is "Stryker Trauma has become aware that five lots of the Gamma3 Set Screwdriver have hexagon tips that were not hardened to specification". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1627-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Gammas Set Screwdriver Instrument Set, Basic Stryker Trauma GmbH, Germany; Distributed by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 - Orthopedic Device. The Gamma3 Locking Nail Systems are intramedullary short and long fixation nails intended for control of rotation and shortening of fractured femurs. The screwdriver is used to place the set screw in the nail.
Product code
HXX Screwdriver
Reason for recall
Stryker Trauma has become aware that five lots of the Gamma3 Set Screwdriver have hexagon tips that were not hardened to specification.
FDA root cause
Finished device change control
Action
On February 22, 2008 Urgent Product Recall notification letters were sent to Stryker Branches/Agencies, Risk Management Departments at hospitals and surgeons via Federal Express. If you have any questions contact Rita Intorrella at (201) 831-5825 or the Hip Fracture Team at (201) 831-5291.
Quantity in commerce
43 total (single and kits)
Distribution
Class II Recall - Nationwide Distribution.
Product codes and lots
Product number: 1320-6000; Lot code engraved on screwdrivers: K887079 and K782153.