AxSYM Ultrasensitive hTSH II Master Calibrators Microparticle Enzyme Immunoassay that determines amount of...
Abbott Diagnostic International, Ltd. began a recall of AxSYM Ultrasensitive hTSH II Master Calibrators Microparticle Enzyme Immunoassay that determines amount of human thyroid stimulating hormone in human serum and... on . The reason FDA records is "High Control value out of range after calibration. --AN INCREASED FREQUENCY OF HIGH CONTROL VALUES BEYOND THE UPPER LIMIT OF THE RANGE SPECIFIED BY THE PACKAGE INSERT". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1566-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Diagnostic International, Ltd.
- Product
- AxSYM Ultrasensitive hTSH II Master Calibrators Microparticle Enzyme Immunoassay that determines amount of human thyroid stimulating hormone in human serum and plasma.
- Product code
- JIS Calibrator, Primary
- Reason for recall
- High Control value out of range after calibration. --AN INCREASED FREQUENCY OF HIGH CONTROL VALUES BEYOND THE UPPER LIMIT OF THE RANGE SPECIFIED BY THE PACKAGE INSERT.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm issued Product Recall letters dated February 22, 2008, requesting remaining units of the affected product be discontinue using and destroyed. A Customer Reply Immediate Action Required letter was enclosed to indicate acknowledging receipt of the product recall letter and to provide information on the number of kits destroyed before March 7, 2008 and faxed to 1-800-777-0051.
- Quantity in commerce
- 882 total units
- Distribution
- Worldwide Distribution --- USA including states of AZ, CA, CO, FL, GA, ID, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, WA, WI, WV, and WY, and Puerto Rico, and countries of Canada, Mexico, Brazil, Chile, Germany, Japan, and Hong Kong.
- Product codes and lots
- Lot 50082Q100 Exp Date 02/23/2008, Lot 52369Q100 Exp Date 04/20/2008, Lot # 53620Q100 Exp Date 06/26/2008 and Lot # 57518Q100 Exp Date 09/07/2008
FIRST ARTICLE / WEEKLY