Z-1616-2008

9051 Electrode Adapter; Automated external defibrillator, Used with QUIK-COMBO System; Cardiac Science Corp...

ZOLL Medical Corporation / initiated 2008-01-23 / Terminated / root cause: manufacturing process

Cardiac Science Corporation began a recall of 9051 Electrode Adapter; Automated external defibrillator, Used with QUIK-COMBO System; Cardiac Science Corp, Bothell, WA 98021 on . The reason FDA records is "Adaptor Cable Incorrectly Manufactured: The red and white connector ends of the adapter are switched. Therefore, the cable will not connect to the electrodes as indicated in the instructions". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1616-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZOLL Medical Corporation
Product
9051 Electrode Adapter; Automated external defibrillator, Used with QUIK-COMBO System; Cardiac Science Corp, Bothell, WA 98021.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Adaptor Cable Incorrectly Manufactured: The red and white connector ends of the adapter are switched. Therefore, the cable will not connect to the electrodes as indicated in the instructions.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified by an Urgent Medical Device Recall letter sent on 01/23/2008. The letter instructs users to check each cable of the affected lot to ensure that it properly connects to the electrode. Users were also instructed to return any defective product to the recalling firm for a replacement. For additional information, contact 1-888-466-8686 x2041.
Quantity in commerce
Lot W029275 - 24; Lot W029475 - 38
Distribution
Nationwide Distribution including states of CT, IA, IN, NC, NY, OH, PA, WA, and WI.
Product codes and lots
Lot Numbers: W029275 and W029476.