Philips Medical Systems iSite PACS (Picture Archiving and Communications Systems) version 4.1.17.0...
Philips Healthcare Informatics, Inc. began a recall of Philips Medical Systems iSite PACS (Picture Archiving and Communications Systems) version 4.1.17.0 - Radiology Image management system on . The reason FDA records is "When using the Freehand Region of Interest (ROI) tool, errors may occur in area calculation yielding incorrect results. The firm discovered that a descrepancy exists in the measurement provided by the tool". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1444-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Medical Systems iSite PACS (Picture Archiving and Communications Systems) version 4.1.17.0 - Radiology Image management system.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- When using the Freehand Region of Interest (ROI) tool, errors may occur in area calculation yielding incorrect results. The firm discovered that a descrepancy exists in the measurement provided by the tool.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm issued product notification letter February 20, 2008, to its consignees. The notification alerts customers to the behavior, requires them to post the notification and notify users not to use the Freehand ROI feature until they are upgraded to a corrected software version. Philips will deliver a software upgrade to correct the problem.
- Quantity in commerce
- 2 units
- Distribution
- Worldwide Distribution --- USA including states of MN, CO, and PA, and country of Germany.
- Product codes and lots
- Version 4.1.17.0