Z-1402-2008
Mevatron KDS-2 Medical Linear Accelerator, Material Number: 9822693, with control console software versions...
Siemens Medical Solutions USA, Inc. / initiated 2008-01-11 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of Mevatron KDS-2 Medical Linear Accelerator, Material Number: 9822693, with control console software versions 6.5/7.2, 8.0, 9.0 and 9.1, Manufactured by Siemens... on . The reason FDA records is "Linear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1402-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Mevatron KDS-2 Medical Linear Accelerator, Material Number: 9822693, with control console software versions 6.5/7.2, 8.0, 9.0 and 9.1, Manufactured by Siemens Medical Solutions USA, Inc, Concord, CA
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Linear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- This is a mandatory software upgrade for consignees. The firm's representatives have contacted consignees via Urgent Device Notification letters dated February 14, 2008, notifying them of the new update and making arrangements to install it.
- Quantity in commerce
- 2 units
- Distribution
- Worldwide-USA including states of AL, AK, AZ, AR, CA, CO, DE, DC, FL, GA, IL, IN, KS, KY, LA, MD, ME, MI, MN, MS, MO, MN, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, VT, VA, WA, WV, WI, and WY and countries of Angola, Australia, Austria,Belgium,Brazil,Canada,Chile,Colombia,Costa Rica, Croatia,Cyprus,The Czech Republic,Denmark,DPR of Korea,El Salvador,France,Germany,Greece,Hungary,India,Iran,Ireland,Italy,Japan,Jordan,Malaysia,Mexico, The Netherlands,New Zealand,Norway,P.R. China,Pakistan,Peru,Philippines,Poland,Portugal,Republic of Korea,Romania,Russian Fed.Saudi Arabia,Singapore,Slovakia,South Africa,Spain,Sweden,Taiwan,Thailand,Trinidad,Tobago,Turkey, U.A.E.,Ukraine, The United Kingdom, and Venezuela
- Product codes and lots
- Serial Numbers: 2518 and 2289