Electrode Adaptor Cable Quick-Combo System, Model 9055 The 9055 Electrode Adaptor is intended for use by...
Cardiac Science Corporation began a recall of Electrode Adaptor Cable Quick-Combo System, Model 9055 The 9055 Electrode Adaptor is intended for use by trained personnel who use a Cardiac Science Automated... on . The reason FDA records is "Instructions For Use is not clear about how the AED voice and text prompts will differ when the adapter is used". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1714-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ZOLL Medical Corporation
- Product
- Electrode Adaptor Cable Quick-Combo System, Model 9055 The 9055 Electrode Adaptor is intended for use by trained personnel who use a Cardiac Science Automated External Defibrillator (AED) for defibrillation with Medtronic QUICK-COMBO Electrodes.
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Instructions For Use is not clear about how the AED voice and text prompts will differ when the adapter is used.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- On 1/23/08 the firm issued a letter titled "9055 Electrode Adaptor IFU Update" to consignees sent via certified return receipt requested. Enclosed with the letter is an updated Instructions For Use, part number 70-00180-01, revision C. Consignees were advised to discard their old Instructions For Use. For assistance, contact Cardiac Science at 1-888-466-8686.
- Quantity in commerce
- 63
- Distribution
- Distributed Nationwide US and Canada and Internationally to: Norway and Guam
- Product codes and lots
- Lot numbers: 123004, 92804, 62305, 63005, 52504, 50405, 33105, 12605, 12505, 7715, 6952, and 6838