Z-1714-2008

Electrode Adaptor Cable Quick-Combo System, Model 9055 The 9055 Electrode Adaptor is intended for use by...

ZOLL Medical Corporation / initiated 2008-01-23 / Terminated / root cause: labeling and UDI

Cardiac Science Corporation began a recall of Electrode Adaptor Cable Quick-Combo System, Model 9055 The 9055 Electrode Adaptor is intended for use by trained personnel who use a Cardiac Science Automated... on . The reason FDA records is "Instructions For Use is not clear about how the AED voice and text prompts will differ when the adapter is used". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1714-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZOLL Medical Corporation
Product
Electrode Adaptor Cable Quick-Combo System, Model 9055 The 9055 Electrode Adaptor is intended for use by trained personnel who use a Cardiac Science Automated External Defibrillator (AED) for defibrillation with Medtronic QUICK-COMBO Electrodes.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Instructions For Use is not clear about how the AED voice and text prompts will differ when the adapter is used.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
On 1/23/08 the firm issued a letter titled "9055 Electrode Adaptor IFU Update" to consignees sent via certified return receipt requested. Enclosed with the letter is an updated Instructions For Use, part number 70-00180-01, revision C. Consignees were advised to discard their old Instructions For Use. For assistance, contact Cardiac Science at 1-888-466-8686.
Quantity in commerce
63
Distribution
Distributed Nationwide US and Canada and Internationally to: Norway and Guam
Product codes and lots
Lot numbers: 123004, 92804, 62305, 63005, 52504, 50405, 33105, 12605, 12505, 7715, 6952, and 6838