Z-1658-2008

Biomet Hybrid Glenoid Boss Cutter; item 406150. - Surgical instrument used to prepare the natural shoulder...

Zimmer GmbH / initiated 2008-03-13 / Terminated / root cause: design

Biomet, Inc. began a recall of Biomet Hybrid Glenoid Boss Cutter; item 406150. - Surgical instrument used to prepare the natural shoulder glenoid surface for implanting a Hybrid Glenoid... on . The reason FDA records is "The glenoid implant may not seat fully on the face of the natural glenoid due to inadequate measurement tolerances (tolerance stack-up) in the design of the cutter. Use of the instrument may lead to incomplete seating...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1658-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet Hybrid Glenoid Boss Cutter; item 406150. - Surgical instrument used to prepare the natural shoulder glenoid surface for implanting a Hybrid Glenoid component. This instrument is intended to function by drilling a hole into the natural glenoid surface allowing the center boss of glenoid implant to sit within the hole, and the non-articulating surface of the glenoid implant to rest flush onto the natural glenoid surface.
Product code
KWT Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented
Reason for recall
The glenoid implant may not seat fully on the face of the natural glenoid due to inadequate measurement tolerances (tolerance stack-up) in the design of the cutter. Use of the instrument may lead to incomplete seating of the glenoid implant, elevating the theoretical risk of glenoid loosening, which may lead to early revision surgery. The problem may contribute to early device failure and patien
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were notified by Urgent Medical Device Removal Notice dated 3/13/08 to discontinue use of the instruments and return to the firm. Operating room managers were sent a separate letter dated 3/13/08 and informed of the possible effects to patients from previous use of this instrument and it was suggested they monitor patients for these effects. If you have any questions contact Mary Hardesty at 1-800-348-9500 or (574) 372-3983.
Quantity in commerce
189
Distribution
Class II Recall - Nationwide Distribution.
Product codes and lots
Lots 44550, 110100, 171760, 211530, 211550, 767940, 820310, 820330, 827250, 853810, 866590, 919780, 919790, 919800, 968270, 980230 and 980250.