Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/REF 00-2550-003-10; 400 ml INF...
Zimmer Inc. began a recall of Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/REF 00-2550-003-10; 400 ml INF. Cntrl. Kit w/ 3/16" drain, 4.8 mm, 15Fr, PVC, 107... on . The reason FDA records is "Leakage: The kits may disassemble at the fluid collection port, which would present a risk of exposure to blood borne pathogens to health care providers should it occur during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1450-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/REF 00-2550-003-10; 400 ml INF. Cntrl. Kit w/ 3/16" drain, 4.8 mm, 15Fr, PVC, 107 cm, soft, sterile, latex free, Rx only; Sold in packages of 10, which contain packages of 1 and 5; Zimmer U.K. Ltd., Swindon, Wiltshire, United Kingdom; Zimmer, Dover, Ohio.
- Product code
- GCY Apparatus, Suction, Single Patient Use, Portable, Nonpowered
- Reason for recall
- Leakage: The kits may disassemble at the fluid collection port, which would present a risk of exposure to blood borne pathogens to health care providers should it occur during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified via Urgent Infection Control Recall letter dated 1/31/08 to examine their inventory and return the recalled lots. Custom kit packers were instructed in the same letter to recall these products/lots from their own customers.
- Quantity in commerce
- 34,354 for (Recalls Z-1448/1458-2008)
- Distribution
- Worldwide-USA, Australia, Canada, Colombia, Denmark, Dubai, Italy, Japan, Mexico and Singapore.
- Product codes and lots
- Lot Numbers: 60006897, 60030744, 60051038, 60067688, 60075523, 60099441, 60113095, 60134967, 60152075, 60168775, 60184523, 60200854, 60227354, 60240752, 60275128, 60299770, 60329438, 60339077, 60367783, 60385889, 60402652, 60437089, 60493507, 60525201, 60554806, 60581694, 60617399 and 60629984.