Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/ REF 00-2550-050-10; 400 ml...
Zimmer Inc. began a recall of Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/ REF 00-2550-050-10; 400 ml Infection Control Comp. Kit, sterile, latex free, Rx... on . The reason FDA records is "Leakage; The kits may disassemble at the fluid collection port, which would present a risk of exposure to blood borne pathogens to health care providers should it occur during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1452-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/ REF 00-2550-050-10; 400 ml Infection Control Comp. Kit, sterile, latex free, Rx only; Sold in packages of 10, which contain packages of 1 and 5; Zimmer U.K. Ltd., Swindon, Wiltshire, United Kingdom; Zimmer, Dover, Ohio.
- Product code
- GCY Apparatus, Suction, Single Patient Use, Portable, Nonpowered
- Reason for recall
- Leakage; The kits may disassemble at the fluid collection port, which would present a risk of exposure to blood borne pathogens to health care providers should it occur during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified via Urgent Infection Control Recall letter dated 1/31/08 to examine their inventory and return the recalled lots. Custom kit packers were instructed in the same letter to recall these products/lots from their own customers.
- Quantity in commerce
- 34,354 (Recalls Z-1448/1458-2008)
- Distribution
- Worldwide-USA, Australia, Canada, Colombia, Denmark, Dubai, Italy, Japan, Mexico and Singapore.
- Product codes and lots
- Lot Numbers: 60006819, 60015774, 60033386, 60045263, 60057329, 60067689, 60084095, 60094656, 60106763, 60116383, 60126841, 60145039, 60158789, 60171589, 60184524, 60200855, 60215687, 60230182, 60240753, 60262574, 60278454, 60294877, 60321541, 60333502, 60342988, 60354480, 60371375, 60385890, 60402653, 60412505, 60437090, 60471395, 60506574, 60520507, 60551036, 60570976, 60590576, 60612123 and 60626622.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.