Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/ REF 00-2550-702-10; 400 ml INF...
Zimmer Inc. began a recall of Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/ REF 00-2550-702-10; 400 ml INF. CNTRL. Kit W/ 1/8" Drain Spec., 3.2mm, 10Fr., PVC... on . The reason FDA records is "Leakage: The kits may disassemble at the fluid collection port, which would present a risk of exposure to blood borne pathogens to health care providers should it occur during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1453-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/ REF 00-2550-702-10; 400 ml INF. CNTRL. Kit W/ 1/8" Drain Spec., 3.2mm, 10Fr., PVC, 107cm, medium, sterile, latex free, Rx only; Sold in packages of 10, which contain packages of 1 and 5; Zimmer U.K. Ltd., Swindon, Wiltshire, United Kingdom; Zimmer, Dover, Ohio.
- Product code
- GCY Apparatus, Suction, Single Patient Use, Portable, Nonpowered
- Reason for recall
- Leakage: The kits may disassemble at the fluid collection port, which would present a risk of exposure to blood borne pathogens to health care providers should it occur during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified via Urgent Infection Control Recall letter dated 1/31/08 to examine their inventory and return the recalled lots. Custom kit packers were instructed in the same letter to recall these products/lots from their own customers.
- Quantity in commerce
- 34,354 (Recalls Z-1448/1458-2008)
- Distribution
- Worldwide-USA, Australia, Canada, Colombia, Denmark, Dubai, Italy, Japan, Mexico and Singapore.
- Product codes and lots
- Lot Numbers; 60004885, 60018681, 60030745, 60045264, 60056512, 60090134, 60101563, 60124258, 60152076, 60189147, 60192877, 60198414, 60217797, 60236478, 60255679, 60281660, 60314033, 60377257, 60402970, 60406391, 60419371, 60444465, 60463238, 60479892, 60511109, 60554807, 60575194, 60586769, 60607929 and 60621372.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.