Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/REF 00-2550-703-10; 400 ml INF...
Zimmer Inc. began a recall of Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/REF 00-2550-703-10; 400 ml INF. CNTRL. Kit W/ 3/16" Drain Spec., 4.8mm, 15Fr., PVC... on . The reason FDA records is "Leakage: The kits may disassemble at the fluid collection port, which would present a risk of exposure to blood borne pathogens to health care providers should it occur during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1454-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/REF 00-2550-703-10; 400 ml INF. CNTRL. Kit W/ 3/16" Drain Spec., 4.8mm, 15Fr., PVC, 107cm, soft, sterile, latex free, Rx only; Sold in packages of 10, which contain packages of 1 and 5; Zimmer U.K. Ltd., Swindon, Wiltshire, United Kingdom; Zimmer, Dover, Ohio.
- Product code
- GCY Apparatus, Suction, Single Patient Use, Portable, Nonpowered
- Reason for recall
- Leakage: The kits may disassemble at the fluid collection port, which would present a risk of exposure to blood borne pathogens to health care providers should it occur during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified via Urgent Infection Control Recall letter dated 1/31/08 to examine their inventory and return the recalled lots. Custom kit packers were instructed in the same letter to recall these products/lots from their own customers.
- Quantity in commerce
- 34,354 (Recalls Z-1448/1458-2008)
- Distribution
- Worldwide-USA, Australia, Canada, Colombia, Denmark, Dubai, Italy, Japan, Mexico and Singapore.
- Product codes and lots
- Lot Numbers: 60045265, 60101564, 60154830, 60196965, 60333503, 60412506, 60453016 and 60633939.