Z-1659-2008

LS-232 Large 12 LeadWear Single Pack - Disposable ECG Monitoring component intended for use as a...

LifeSync Corporation / initiated 2007-12-13 / Terminated / root cause: manufacturing process

Lifesync Corporation began a recall of LS-232 Large 12 LeadWear Single Pack - Disposable ECG Monitoring component intended for use as a radiofrequency signal transmitter and receiver of... on . The reason FDA records is "LifeSync Wireless ECG System - Intermittent failure when using LS-202 and LS-232 LeadWear in conjunction with LS-41245 and LS-41285 Adaptors. The out of box failure manifests as a V-Lead failure, leading to a possible...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1659-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LifeSync Corporation
Product
LS-232 Large 12 LeadWear Single Pack - Disposable ECG Monitoring component intended for use as a radiofrequency signal transmitter and receiver of electrocardiograph physiological signals which are displayed on the ECG monitors of various manufacturers' systems that have been validated for compatibility.
Product code
DRG Transmitters And Receivers, Physiological Signal, Radiofrequency
Reason for recall
LifeSync Wireless ECG System - Intermittent failure when using LS-202 and LS-232 LeadWear in conjunction with LS-41245 and LS-41285 Adaptors. The out of box failure manifests as a V-Lead failure, leading to a possible delay in treatment based upon a delay in ECG monitoring.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Firm issued a Device Recall Letter on 12/14/2007. Users were requested to remove the product from inventory and discontinue use of the Large Sized LeadWear with the 12 Lead Adaptors. The current stock of product in inventory will be held under a stop shipment, pending destruction of non-conforming product.
Quantity in commerce
768 completed leadwire units
Distribution
Class II Recall - Nationwide Distribution.
Product codes and lots
All lots