LS-202 Large LeadWear Combo Pack - Disposable ECG Monitoring component intended for use as a radiofrequency...
Lifesync Corporation began a recall of LS-202 Large LeadWear Combo Pack - Disposable ECG Monitoring component intended for use as a radiofrequency signal transmitter and receiver of... on . The reason FDA records is "LifeSync Wireless ECG System - Intermittent failure when using LS-202 and LS-232 LeadWear in conjunction with LS-41245 and LS-41285 Adaptors. The out of box failure manifests as a V-Lead failure, intended for use as a...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1660-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LifeSync Corporation
- Product
- LS-202 Large LeadWear Combo Pack - Disposable ECG Monitoring component intended for use as a radiofrequency signal transmitter and receiver of electrocardiograph physiological signals which are displayed on the ECG monitors of various manufacturers' systems that have been validated for compatibility.
- Product code
- DRG Transmitters And Receivers, Physiological Signal, Radiofrequency
- Reason for recall
- LifeSync Wireless ECG System - Intermittent failure when using LS-202 and LS-232 LeadWear in conjunction with LS-41245 and LS-41285 Adaptors. The out of box failure manifests as a V-Lead failure, intended for use as a radiofrequency signal transmitter and receiver of electrocardiograph physiological signals which are displayed on the ECG monitors of various manufacturers' systems that have been
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Firm issued a Device Recall Letter on 12/14/2007. Users were requested to remove the product from inventory and discontinue use of the Large Sized LeadWear with the 12 Lead Adaptors. The current stock of product in inventory will be held under a stop shipment, pending destruction of non-conforming product.
- Quantity in commerce
- 576 completed leadwire units
- Distribution
- Class II Recall - Nationwide Distribution.
- Product codes and lots
- All Lots